Method basis: National standard GB11894-89 "Water quality - Determination of total nitrogen - UV spectrophotometric method》.
2 Measurement range:0100mg/LTotal nitrogen (graded)0-1mg/L;0.1-30mg/L;1-50mg/L; 1-100mg/L).
3 Accuracy:Not exceeding±10%Or not exceeding±0.2mg/L.
4 Repeatability:Not exceeding±5%Or not exceeding±0.2mg/L.
5 measurement cycles: The minimum measurement cycle is 20 minutes, and according to the actual water sample, the digestion time can be arbitrarily modified within 5-120 minutes.
6 sampling periods: time interval (adjustable from 10 to 9999 minutes) and whole point measurement mode.
Calibration cycle: 1 to 99 days, adjustable at any interval and time.
8 Maintenance cycle: Generally once a month, about 30 minutes each time.
9. Reagent consumption: less than 0.5 yuan/sample.
10 outputs: 2roadRS-232。1road 4~20mA
11 Environmental requirements: For indoor environments with adjustable temperature, it is recommended to have a temperature range of+5-28 ℃; Humidity ≤ 90% (non condensing).
12 Power supply: AC230 ± 10% V, 50 ± 10% Hz, 5A.
Size 13: Height 1450 x Width 510 x Depth 450mm.
14 Other: Abnormal alarms and power outages will not result in data loss;
Touch screen display and command input;
After abnormal reset and power failure, the instrument will automatically discharge residual reactants and restore working status.