ZD 2000typeThe fully automatic oscillation instrument is mainly used for oscillation degassing in the determination of dissolved gas components in oil by gas chromatography method; The oscillation degassing method for determining water-soluble acids in oil (colorimetric method) can also be used for constant temperature and timed oscillation in other physical and chemical experiments. It adopts32-bit embeddedminutetypecomputercontrol,LCD display technology, Chinese menu display, more convenient to use and operate. Temperature control adopts microcomputer fuzzy calculation for temperature controlreachPID, etcThe method and temperature control accuracy have been further improved. The sliding door design makes it easier to pick up and place samples. It has the characteristics of small size, low noise, uniform temperature field distribution, and eight samples per oscillation. The fully automated work process reduces the labor intensity of workers and improves work efficiency, making it an ideal replacement product for laboratories.
Main technical parameters
Applicable standards:GB/T 7601-87
Temperature control range: room temperature~100ºC
Temperature control accuracy: room temperature~60ºC±0.3ºC
60~100ºC±0.5ºC
Oscillation frequency:275 ± 3 times/minute
Oscillation amplitude:35 millimeters
Number of samples per oscillation:
8 syringes of 50 or 100 milliliters;
5 250ml triangular flasks;
Power supplyvoltage:AC220V±10%
Frequency:50Hz±5%
Power:900VA
External dimensions(LxWxH):450×330×410
Main features
■ Time control methodSet the oscillation time and static time arbitrarily within 0-99 minutes
Noise level less than 35 decibels
■ Temperature control methods: microcomputer fuzzy control FUZZY, PID, SELF and other temperature control methods