XY-21BS double single-sided standard horizontal purification table
maintenance
Product details

Installation and use of XY-JH-11BS_XY-JH-21BS purification workbench

The workstation should be located in a relatively clean working environment (preferably in a primary purification room of 100000 or 300000 levels), plugged in the power, and turned on according to the functions shown on the control panel. Before starting up, the working area and shell of the purification workstation should be thoroughly cleaned to remove surface dust. Normal operation and use can be carried out within ten minutes after starting up.

maintenance

1. According to the actual situation, the primary filter should be removed and cleaned regularly, with a cleaning cycle of generally 3-6 months. If not cleaned for a long time, dust accumulation will lead to insufficient air intake and reduce the cleaning effect.

2. When the ideal cross-sectional wind speed cannot be achieved after normal replacement or cleaning of the primary air filter, the operating voltage of the fan should be adjusted to achieve the ideal uniform wind speed.

3. Generally, when the operating voltage of the fan is adjusted to the highest point after 18 months of use and still cannot reach the ideal wind speed, it indicates that the high-efficiency air filter has accumulated too much dust (the filter holes on the filter material are basically blocked and need to be updated in a timely manner). The service life of a high-efficiency air filter is generally 18 months.

Online inquiry