XA-100 Comprehensive Atmospheric Sampler (Electronic Heating)
XA-100 Comprehensive Atmospheric Sampler (Electronic Heating) I. Implementation Standard JJG956-2000 "Atmospheric Sampler" JJG943-2011 "Total Suspende
Product details

XA-100 Comprehensive Atmospheric Sampler (Electronic Heating)
1、 Execution standards
JJG956-2000 "Atmospheric Sampler" JJG943-2011 "Total Suspended Particle Sampler"
HJ/T374-2007 Technical Requirements and Testing Methods for Total Suspended Particle Samplers
HJ/T375-2007 "Technical Requirements and Testing Methods for Environmental Air Samplers"
HJ618-2011 "Determination of PM10 and PM2.5 in ambient air by weight method"
HJ/T93-2003 Technical Requirements and Testing Methods for PM10 Samplers
2、 Purpose:
The application of the national standard solution absorption method to collect toxic and harmful gases in the ambient air, and the national standard filter membrane weight method to collect particulate matter in the ambient air, is a new type of comprehensive atmospheric sampler that integrates the collection of harmful gases and particulate matter. It is suitable for unmanned all-weather work and is equipped with a high-strength aluminum alloy PM2.5 multi-stage particulate cutter to collect particles such as TSP, PM10, PM2.5, PM1, etc. It can be widely used in environmental monitoring, health and epidemic prevention, labor protection, scientific research institutions and other departments for sampling harmful gases and particulate matter.
3、 Performance characteristics
1. Featuring small size, light weight, low noise, high level of intelligence, stable flow rate, reliable operation, built-in heater, illustrated text, and integrated collection of harmful gases and particulate matter from the atmosphere.
2. Adopting imported wide temperature OLED display screen, with a good human-machine interface and vivid graphics, easy operation, and user-friendly Chinese graphic display, it can work in ultra-low temperature environments.
3. The microcomputer simultaneously controls three sets of sampling systems for two atmospheric channels and one particulate matter channel, with multiple sampling methods including simultaneous sampling, individual sampling, multiple sampling, alternate day sampling, and cyclic sampling.
4. The dual channel atmospheric and particulate matter three-way sampling systems both use imported DC brushless vacuum pumps, with a fault free usage time of over 20000 hours.
5. The instrument is a dual-purpose AC/DC type, equipped with a built-in large capacity imported lithium battery cell. It can work in situations without AC power supply and has a continuous working time of more than 3 hours (when all three channels are fully open and working continuously).
6. Adopting imported pressure and flow sensors, the sampling flow accuracy is high and the stability is good. It can measure atmospheric pressure and temperature, and automatically convert states based on pressure and temperature.
7. Parameters such as sampling flow rate, startup time, sampling time, interval time, and number of cycles can be set.
8. Automatic fault protection function, users can set their own limit protection flow pressure, and if the set flow rate is not reached within the specified time, automatic shutdown protection will be implemented to prevent damage.
9. Built in rechargeable battery for storing data during AC power outages, automatically restoring sampling during AC incoming calls, deducting the power outage time during the sampling process, and allowing users to query the power outage time.
10. The unique three in one design of drying, filtering, and anti backflow suction in the drying cylinder can efficiently dry moisture, filter dust, and reduce the impact of backflow suction on flow rate, achieving long-term operation without cleaning.
11. Users can calibrate various parameters of the instrument through the waterproof keyboard, but they need to enter with a password to ensure the security of the instrument data.
12. Sampling settings and automatic memory settings for sampling data, priority should be given to the next startup, achieving one click sampling.
13. The XA-100 comprehensive atmospheric sampler has a built-in heating device that can set the heating temperature and display the temperature of the absorption bottle. It can perform constant temperature sampling in low-temperature environments.
14. The tripod cloud desktop design is convenient for users to use, and the XA series cutters are made of high-strength aluminum alloy material with anti-static adsorption.
15. Data printing can be done by selecting a micro printer through the RS232 interface, or it can be exported through a USB flash drive.
4、 Technical parameters

Parameters

Parameter range

resolution

accuracy

air sampler

Dual electronic flow meter (0.1-1.0) L/min

0.1L/min

≤±2.5%

Dust sampling flow rate

(50~150)L/min

0.1L/min

≤±2.5%

Pre measurement pressure

(-30~0)Kpa

0.01 Kpa

≤±2.5 %

Effective diameter of filter membrane

Particle filter membrane with a diameter of 90 mm and an effective area of 80 mm

Maximum sampling volume

999999.99L

control mode

Automatic or manual, set the on/off time arbitrarily within 24 hours

Delay Time

1min~99h59min

1min

≤±2S

sample time

1min~99h59min

1min

≤±2S

Timed startup

1min~99h59min

1min

≤±2S

Interval time

1min~99h59min

1min

≤±2S

atmospheric pressure

(70~130)Kpa

0.1Kpa

≤±2.5%

Traffic stability

≤ 2% (with resistance variation of 3 Kpa~6Kpa and grid fluctuation of ≤± 10%)

Traffic response

<5S

Timing accuracy

24 hours ± 2S

Sampling times

1 to 99 times

Power outage protection time

More than 5 years

Pre measurement temperature

(-30~+99)℃

1℃

≤±2.0%

Heating temperature control range

(10-30) ℃, factory set at 23 ℃

0.1℃

≤±1%

Working power supply

AC220V ± 10% 50Hz or built-in large capacity lithium battery (optional)

noise

<50dB(A)

power consumption

<20W (excluding heater)

External dimensions

(Length 255 x Width 265 x Height 185) mm

Whole machine weight

About 6kg