Technical parameters of walk-in low-pressure laboratory and walk-in high and low temperature low-pressure laboratory
1、 The high-altitude test shall be conducted according to the following steps:
- The test object is a battery pack or system.
- Test environment: Atmospheric pressure conditions at an altitude of 4000m or equivalent, with a temperature of room temperature.
- Leave the test object in the testing environment of 7.12.2 for 5 hours and perform a constant current discharge of 1C (not exceeding 400A) until the discharge cut-off condition is reached. Observe for 2 hours.
- Requirement: The battery pack or system should have no sharp changes in discharge current, abnormal voltage, leakage, shell rupture, fire or explosion, etc. The insulation resistance value after the test shall not be less than 100 Ω/V
2、 Main technical parameters
instrument for drawing circles |
model |
CZ-GQ-8000E |
CZ-GQ-16000E |
CZ-GQ-24000E |
CZ-GQ-120000E |
internal volume |
8000L |
16000L |
24000L |
120000L |
|
Studio Size |
W2000×H2000×D2000Mm |
W2000×H2000×D4000mm |
W3000×H2000×D4000mm |
W8000×H2500×D6000mm |
|
External dimensions |
Based on actual dimensions |
Based on actual dimensions |
Based on actual dimensions |
Based on actual dimensions |
|
External material |
Thickened cold-rolled steel plate, treated with baking paint (milky white) |
||||
Inner box material |
Thickened SUS # 304 stainless steel for easy cleaning |
||||
Box door |
Thickened cold-rolled steel plate, baked paint (blue) treatment, double door |
||||
Observation window |
Tempered glass observation window |
||||
pressure relief device |
Box configuration with pressure relief port |
||||
host |
Control method |
Touchscreen LCD color display screen |
|||
temperature range |
-50~+100 ℃ (customizable) |
||||
cooling rate |
+25 ℃ → -50 ℃ 1 ℃/min (average throughout the entire process) |
||||
heating rate |
+25 ℃ →+100 ℃ 3 ℃/min (average throughout the entire process) |
||||
Temperature resolution |
0.1℃; |
||||
temperature fluctuation |
±0.5℃; (Constant state under normal pressure and no load) |
||||
Temperature deviation |
±2.0℃; (Constant state under normal pressure and no load) |
||||
temperature uniformity |
≤2.0℃; (Constant state under normal pressure and no load) |
||||
Pressure Range |
Normal pressure~10kPa; |
||||
Voltage reduction time |
Normal pressure~10kPa ≤ 5min (under -50 ℃ environment) |
||||
Air pressure change rate |
≥10kPa/min |
||||
Pressure deviation |
Normal pressure~40KPa: ≤± 2KPa; |
||||
40KPa~2KPa:≤±5%; |
|||||
2KPa~1kpa: ≤±0.1KPa |
|||||
Camera monitoring system |
Configure high-definition camera system (optional) within the testing area |
||||
Standard configuration |
2 sets of host, sample rack, and sample vehicle |
||||
Auxiliary functions |
Voltage and temperature collection, safety and explosion-proof system, smoke emission system, smoke filtration system (optional) |
||||
Safety protection |
Leakage, overtemperature, overvoltage, overload, abnormal sound and light alarm |
||||
supply voltage |
AC380V 50Hz |