Qingdao Hengyuan Technology Development Co., Ltd
Home>Products>Ultra large flow dioxin sampler
Ultra large flow dioxin sampler
Purpose: Used for collecting semi volatile organic compounds (SVOCs) samples such as polycyclic aromatic hydrocarbons, organic pesticides, and polychl
Product details

purpose

Used for collecting semi volatile organic compounds (SVOCs) samples such as polycyclic aromatic hydrocarbons, organic pesticides, and polychlorinated biphenyls in the ambient air


Adopting standards

HJ 691-2014 "Guidelines for Sampling Techniques of Semi volatile Organic Compounds in Environmental Air"

HJ 902-2017 "Determination of Polychlorinated Biphenyls in Environmental Air - Gas Chromatography Mass Spectrometry Method"

HJ 903-2017 "Determination of Polychlorinated Biphenyls in Environmental Air - Gas Chromatography Method"


main features

1. Using flow, pressure, temperature sensors, and microcontroller measurement and control technology to automatically control the flow of the vacuum pump and achieve constant flow sampling

2. Automatically measure environmental temperature, atmospheric pressure, flow rate and other parameters

3. Digitally measure parameters such as sampling flow rate (operating conditions), sampling time, interval sampling time, sampling volume (standard conditions), etc

4. Digitally calibrate flow, temperature, pressure, time, and other numerical values

5. Automatically display values such as flow rate, sampling volume (standard condition 237.15k, 101.325kPa), temperature, pressure, sampling time, power outage time, etc

6. Overload automatic protection: During the sampling process, if the sampling flow rate is less than 90% of the set sampling flow rate for 5 consecutive minutes, the instrument will automatically shut down for protection

7. Power outage protection function: During the sampling process, if there is a power outage, the sampling data can be automatically saved, and sampling can continue after the power is restored

8. Can store 100 sets of sampling data and power outage data, with a retention period of 10 years

9. Adopting a brushless fan, it has a long service life, can run for 20000 hours without faults, and has low noise


TECHNICAL INDEX

1. Sampling flow rate: 800L/min; Accuracy: better than 2%; Stability: better than 2% (SVOCs sampling)

2. Timing accuracy: better than 0.1%

3. Sampling flow rate setting range: (200~1200) L/min (up to 1200L/min for particulate matter and benzo [a] pyrene sampling)

4. Temperature measurement range: (-55~150) ℃; Resolution: 0.1 ℃; Accuracy: 0.3 ℃

5. Pressure measurement range: (50~140) kPa; Resolution: 0.01kPa; Accuracy: better than 2 ‰

6. Load capacity: able to overcome 14kPa resistance at 800L/min

7. Sampling method:

Quantitative sampling: Sampling volume setting range (1~9999999.9) L (standard condition)

Immediate sampling: Sampling time setting range (00h01min~24h00min)

Timed sampling: Set the starting time range (00:00~23:59) and sampling time range (00h01min~24h00min)

Interval sampling: 99 interval samples can be set within 24 hours

8. Working voltage: AC (220 ± 22) V, 50Hz


Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!