Toxic and harmful gas detection sensor BYG511
Toxic and harmful gas detection sensor BYG511
Product details
1、 Introduction to Toxic and Harmful Gas Detection Sensors
The BYG511 series toxic and harmful gas detection sensor adopts standard 4 series/7 series electrochemical gas sensors, infrared gas sensors, catalytic combustion gas sensors, output signal TTL, RS485, etc. It is designed for low-cost application users of integrators. As long as there are 4 series/7 series sensors, they can be calculated and calibrated through this module. The existing CO, O3, SO2, CH2O, O2, CxHx sensors have been widely used in various monitoring and monitoring solutions. The intrinsic safety design, hot swappable, simplified peripheral circuit, personalized data calibration, calibration program, sensor probe can be freely switched (of the same model, so that the sensor life can be freely replaced when the limit is reached), and it also supports oxygen cells, 2pin The electrode and similar packaging are perfectly compatible with standard inner diameter explosion-proof detection heads, suitable for rapid prototyping monitoring solutions and environmental gas monitoring systems.
2、 Application fields of toxic and harmful gas detection sensors
This product is widely used in scientific research institutions, portable/handheld detectors, fixed gas transmitters, atmospheric environment detectors, industrial toxic and harmful gas transmitters, etc.
3、 Technical parameters of toxic and harmful gas detection sensors
3、 Technical parameters of toxic and harmful gas detection sensors
| Technical Parameter | BYG511 series |
| Detect gas | 4 or 7 series sensors for toxic and harmful gases, flammable and explosive gases, oxygen, TVOC, etc |
| Detection Principle | Electrochemical, catalytic combustion, infrared, thermal conductivity, PID photoionization, etc., depending on gas, range, on-site environment, and user requirements. |
| measuring range | According to the parameters (refer to the table below), the range can be customized |
| resolution | 0.1/0.01/0.001ppm |
| accuracy | ≤± 3% F.S (higher precision can be customized) |
| Sensor lifespan | Electrochemistry takes 2-3 years, oxygen takes 2 years, infrared principle takes 5-10 years, catalytic combustion takes 3 years, and thermal conductivity takes 5 years. |
| Warm-up time | <3 minutes |
| sampling interval | 3S |
| response time(T90) | <60S |
| Nonlinear | ±2% |
| repeatability | ±2% |
| Supply Voltage | DC5V |
| signal output | TTL、RS485 |
| work environment | -20~50℃; 0-95% RH (no condensation) |
| Environment | -40~60℃; 0-100% RH (no condensation) |
| Weight/Size | φ37*25mm |
| notes | |
4、 Parameter selection table for toxic and harmful gas detection sensors
| Detect gas | model | Detection range (standard) | accuracy | Resolution (standard) |
| oxygen | BYG000-O2 | 0-30% | ±1%FS | 0.1% |
| carbon monoxide | BYG000-CO | 0-1000ppm | ±3%FS | 1ppm |
| hydrogen sulfide | BYG000-H2S | 0-20ppm | ±5%FS | 0.01ppm |
| nitrogen dioxide | BYG000-NO2 | 0-200ppm | ±5%FS | 0.01ppm |
| nitric oxide | BYG000-NO | 0-10ppm | ±5%FS | 0.01ppm |
| sulfur dioxide | BYG000-SO2 | 0-10ppm | ±3%FS | 0.01ppm |
| ammonia | BYG000-NH3 | 0-20ppm | ±3%FS | 0.1ppm |
| ozone | BYG000-O3 | 0-20ppm | ±5%FS | 0.01ppm |
| hydrogen cyanide | BYG000-HCN | 0-200ppm | ±5%FS | 0.1ppm |
| formaldehyde | BYG000-HCHO | 0-2ppm | ±5%FS | 0.01ppm |
| chlorine | BYG000-Cl2 | 0-200ppm | ±5%FS | 0.1ppm |
| Combustible gas | BYG000-EX | 0-100%LEL | ±5%FS | 0.1%LEL |
| hydrogen | BYG000-H2 | 0-2000ppm | ±3%FS | 1ppm |