Product Introduction:
The toxic and harmful substances in indoor air will damage the indoor environment of civil construction projects and pollute human health. The "Control Specification" identifies radon, formaldehyde, ammonia benzene, volatile organic compounds (TVOC), and free toluene diisocyanate (TDI) as the primary pollutants.
Total volatile organic compounds (TVOC) refer to the total amount of volatile organic compounds with a boiling point lower than or equal to 250 ℃ measured in air under normal pressure conditions. (mainly xylene, etc.).
TengzhouHongxiang Analytical InstrumentThe indoor air pollutant detection (TVOC) analysis solution provided by the limited company is based on the requirements of the national air detection standard GB50325-2001 (2006 edition) and the Ministry of Construction's "Code for Indoor Environmental Pollution Control of Civil Building Engineering", using gas chromatography to quantitatively detect the content of TVOC and benzene.
Indoor Environment (TVOC) Analysisgas chromatographConfiguration plan:
1. Hongxiang GC-2010 gas chromatograph dual FID, packed column injection, capillary injection
2.chromatographic columnSE-30 50m*0.53mmdomestic
3.The thermal analyzer analyzes the temperature at 300 ℃, which is domestically produced
4.Oil free ZK-800 air generator
5.Hydrogen generator drying ZH-300
6 .Nitrogen 99.999%+steel cylinder+pressure reducing valve self provided
7.Domestic chromatography workstation
8.UStainless steel adsorption tube (filled with Tenax adsorbent)
9.UStainless steel adsorption tube (filled with activated carbon)
10.UStainless steel adsorption tube (empty tube)
11.Computer, printer to be provided by oneself
12. Domestic single air path atmospheric sampler
Desorption conditions:
Desorption temperature 250 ℃
Desorption time 5 minutes
Sweep air flow rate 50ml/min
Carrier gas nitrogen
Gas chromatograph analysis conditions:
Vaporization temperature 250 ℃
Detecting temperature of 250 ℃
The column chamber temperature is maintained at 50 ℃ for 10 minutes, and then raised to 250 ℃ at a rate of 5 ℃/min
Chromatographic column 50m × 0.53mm × 1.0um
Indoor Environment (TVOC) Analysis Gas Chromatography Analysis Process:
1) Sampling
2) Analysis
3) Gas chromatograph analysis
4) Chromatography workstation
5) Data output
|
|||
|