Instrument Overview:
The TR-308H COD ammonia nitrogen total phosphorus turbidity analyzer is developed and designed based on authoritative certification methods. It can measure four parameters in water: COD, ammonia nitrogen, total phosphorus, and turbidity. The detection principles of each parameter comply with the recognized methods in the environmental protection industry. Through imported cold light sources and narrowband interference technology, the absorbance of each water sample is detected by spectrophotometry at specific light source wavelengths. After automatic data processing by a microcomputer, the concentration values of COD, ammonia nitrogen, total phosphorus, and turbidity in the sample are directly displayed in units of mg/L or degrees.
Technical parameters:
Measurement parameters | COD | Ammonia nitrogen | total phosphorus | turbidity |
Measurement range | 0-15000mg/L | 0-50mg/L | 0-20mg/L | 0-1000mg/L |
Anti chlorine interference | 1000mg/L | - | - | - |
limit of detection | 5mg/L | 0.01mg/L | 0.02mg/L | 0.5mg/L |
Dissolve the environment | 165℃,15min | No need to dissolve | 125℃,30min | No need to dissolve |
Measurement time | 20-30 minutes | 10-15 minutes | 35-45 minutes | 1 minute |
Measurement Error | ≤± 5% | |||
repeatability | ≤±5% | |||
Optical stability | ≤± 0.001A/20 minutes (100000 hours lifespan) | |||
Colorimetric method | Colorimetric tube (digestion colorimetric integrated tube) | |||
Batch processing volume | 6, 16, 25 (optional) | |||
data processing | 4000 records, 96 curves (supporting user-defined calibration) | |||
transmission mode | USB data cable, LAN, Bluetooth transmission, etc. (optional) | |||
printer | Portable thermal printer (optional) | |||
Operation interface | Chinese/English (optional) | |||
Power supply mode | 5V rechargeable battery | |||
power | 0.3W | |||
Host size | 240mm*95mm*80mm | |||
Instrument weight | Host: 0.5kg; Dissolver: 1.37kg | |||
ambient temperature | 5~40℃ | |||
ambient humidity | ≤ 85% non condensing |
Dissolver parameters:
Instrument Name | Multi functional intelligent digestion instrument | ||
Instrument model | TR-6B (standard) | TDR-16A (optional) | TDR-25 (optional) |
Number of holes in the digestion instrument | 6 holes | 16 holes | 25 holes |
Operation interface | Chinese/English (optional) | Full Chinese display | |
rated voltage | DC 12V/10A | AC 220V±10%/50Hz | |
Power supply mode | Adapter, power bank, car mounted | 220V AC power | |
maximum power | 120W | 650W | |
External dimensions | 185*120*155(mm) | 230*340*130(mm) | |
Instrument weight | 1.37kg | 6kg | |
Dissolve pore size/depth | 16mm/65mm | ||
screen display | tft lcd | ||
Standard digestion mode | COD、 Total phosphorus, total nitrogen, total chromium (optional) digestion mode | ||
alarm mode | buzzer | ||
Custom time | 5-999min adjustable | ||
Custom Temperature | Adjustable from 90 to 200 ℃, with constant temperature control for the target temperature | ||
Temp Accuracy | ±1℃ | ||
heating rate | Room temperature -165 ℃, 12 minutes | ||
ambient temperature | (-10-40)℃ | ||
ambient temperature | Relative humidity<85% (no condensation) |
Performance characteristics:
1. Wide detection range: dozens of detection parameters can be expanded and customized according to needs.
2. High sensitivity: The minimum displayed value can reach 0.001 level.
3. Light source advantage: using imported cold light source, with excellent optical performance, no need for preheating, and a lifespan of up to 100000 hours.
4. Intelligent operation: Intelligent operation program that guides users to easily complete operations.
5. Diversified testing methods: Standard measurement or continuous measurement mode can be selected, freely choosing between speed and accuracy.
6. Low usage cost: low consumable prices and minimal usage.
7. Operation safety: Using a digestion colorimetric integrated tube, the measurement is simple, fast, and safe.
8. Large LCD screen display, direct data reading, simple and time-saving operation.
Advantages of emergency detection:
1. The instrument is designed to be compact, with a box like configuration, lightweight, and easy to carry.
2. Provide specialized testing reagents to minimize user operation as much as possible.
3. Supports wireless Bluetooth printing and can print historical measurement data (optional Bluetooth printer).
4. Can store 4000 sets of historical measurement data.
5. Equipped with a large capacity rechargeable lithium battery and a remaining power display, it has low power and power consumption, achieving ultra long standby. The digestion instrument has three power supply modes for users to choose from: portable power supply, vehicle mounted, and 220V AC power.
Online operation function:
Supports network port, USB interface, Bluetooth function, and can transfer data with PC or printer.
Software advantages:
1. Supports Chinese/English display (optional), designed to be friendly, intelligent, and easy to operate.
2. Built in system program and user program, which can add frequently tested items to the user program for easy searching.
3. It has data storage and power-off protection functions to prevent data loss and facilitate the retrieval of historical measurement data.
4. Built in standard operating procedures, no need for user calibration and calibration.
5. Users can also develop their own user programs and reagents to adapt to different situations and water conditions.
6. Built in help interface, explaining the general precautions for user detection process.
7. Equipped with a one click factory reset function.
Advantages of the digestion instrument:
1. Optional 6-well, 16 well, or 25 well digestion instruments are available,moreTo meet the needs of users to a large extent.
2. Adopting intelligent PID temperature control technology and dual anti overheating protection system, heating is safe, uniform, and fast.
3. Built in standard digestion program, one click digestion design, only need to press the confirm button to complete the entire digestion process.
4. Support custom modes, allowing for free setting of digestion temperature and time, achieving one click digestion in the target mode.
5. Used for the digestion of COD, total phosphorus, total nitrogen, total chromium and other projects.
Testing standard basis:
Measurement parameters | Reference standards | Reference standards |
COD | Determination of Chemical Oxygen Demand in Water Quality - Rapid Digestion Spectrometric Method | HJ/T399-2007 |
Ammonia nitrogen | Water quality - Determination of ammonia nitrogen - Nessler's reagent spectrophotometric method | HJ535-2009 |
total phosphorus | Determination of total phosphorus in water quality - Ammonium molybdate spectrophotometric method | GB11893 |
turbidity | Determination of Turbidity in Water Quality | GB13200-91 |