1. Main specifications:
Display mode: 8-inch color touch screen display
Workstation: Application work interface based on Microsoft Windows operating system
Light source: 6V10W imported 2000 hour long life tungsten halogen lamp, 1000 hour long life deuterium lamp
Sample Room: 100mm measurement path length
Standard colorimetric pool holder: 10mm
Power requirements: 100/240VAC, 47-63Hz, ± 10%
Dimensions: 475 × 420 × 180 (mm)
Net weight of the instrument: 11 kilograms
2. Optical performance:
Optical system: Proportional monitoring dual beam, self timing (Littrow type) optical path, 1200 diffraction gratings per millimeter monochromator
Measuring optical path length: 100mm
Spectral bandwidth: 2nm
Wavelength range: 190-1100nm
Wavelength accuracy: ± 0.5nm
Wavelength repeatability: ≤ 0.2
Luminous accuracy: ± 0.3% T
Luminous repeatability: 0.15% T
Stray light: ≤ 0.05% T