Saccharomat is a fully automatic polarimeter for the sugar industry.
Quartz has almost the same optical dispersion (ORD) as sucrose solution. This principle has been uniquely applied as a quartz wedge compensation mechanism by S+H Schmidt Hanxi company Saccharomat. This working principle is currently a unique patented technology.
Due to the fact that quartz and the sample exhibit offset reactions towards small wavelengths in the same way, these offsets are automatically compensated for, thus the dispersion effect is automatically cancelled out. This ensures long-term stable high-precision measurement results without the need for recalibration during the instrument's lifespan. The biggest advantage is that even samples with very dark colors and high absorption (light) ability can be directly measured for optical rotation using NIR near-infrared wavelengths after filtration (using AutoFit automatic filter) without the need for additional clarification.
The quartz wedge automatic polarimeter Saccharomat can be directly combined with the refractometer measurement unit to form a complete automatic purity analyzer, which can simultaneously measure the optical rotation (international standard sugar content) and Brix value, thus directly obtaining the purity value. This combination scheme is extremely economical and efficient.
characteristic:
The fully automatic polarimeter (sugar meter) adopts a unique quartz wedge compensation principle
Compared with a circular polarimeter, its wavelength dependence is smaller
The instrument does not need to be recalibrated at any time
Samples with darker colors can be directly measured after being filtered using Autofilt Z
Can use near-infrared wavelength (NIR)
The instrument has high precision and stability throughout the entire measurement range
Extremely fast temperature and wavelength compensation speed
High photosensitivity
By selecting an appropriate polarizing tube, it can have efficient automatic temperature compensation
Optical rotation tubes with self sealing connection design can be selected for safe and fast plugging and unplugging of water bath interfaces
Triple Quartz Control Panel (TQCP) optional
Users can customize different testing methods
Up to 1000 different user ranges can be set (display units)
CONTINUE
High resolution 7 "TFT touch screen display, user-friendly human-machine interaction operating system
Almost maintenance free
Compliant with GLP/GMP standards
High quality 'Made in Germany', providing online consultation and services.
Software features:
Clear menu guidance for operation
Independence of display and data output
Password protected graded login system (user management)
Freely programmable formulas
Can export data to MS Excel to generate text files or PDF files
Easy to connect to LIMS system (Laboratory Integrated Management System)
Easily reset all data in the original state of the instrument
Easily reset specific settings based on administrator permission levels
Self diagnosis and graded maintenance functions, easier to maintain and serve
Remote diagnosis and data maintenance
User range ° Z international standard sugar content range -35 - 105°Z accuracy 0,01°Z accuracy ± 0,01°Z** repeatability 0,01°Z light sensitivity Up to OD 5 Time About 4 seconds wavelength Standard configuration: 1 or 2 sets of wavelengths Measurement mode CONTINUE temperature Measuring temperature Adapt the temperature sensor based on the rotation tube Sample temperature control External circulating water bath temperature control Temperature range Water bath temperature control: varies according to different circulating water baths Temperature Accuracy Different according to different circulating water baths/0.01 ° C Temperature accuracy Different according to different circulating water baths/± 0.03 ° C more parameters
polarimeter tube Up to 200mm in length ambient temperature +10 - +40 °C light source LED display 7 "TFT touch screen, 800 x 480 pixels, 16 bit color operate Touch screen, keyboard *, mouse *, barcode scanner reader *, PC online* Interface/Online RS232 (1x), USB A (4x), USB B (1x), Ethernet (1x), analog, W-LAN/LAN* power consumption 90 V / 50 Hz - 31,0 W model Saccharomat ® 103 T (587 nm)Main specifications
Alternatively, T-Cell Peltier independent temperature controlled optical tube can be used
Or use Peltier independent temperature controlled rotary tube
Peltier independent temperature controlled rotary tube: 18-40 ° C
90 V / 60 Hz - 31,1 W
110 V / 50 Hz - 31,0 W
110 V / 60 Hz - 30,9 W
230 V / 50 Hz - 33,8 W
230 V / 60 Hz - 33,2 W
standard model
Saccharomat ® 101 T (882,6 nm)
Saccharomat ® 202 T (587 and 882,6 nm)
Main specifications
User range |
° Z international standard sugar content |
range |
-35 - 105°Z |
accuracy |
0,01°Z |
accuracy |
± 0,01°Z** |
repeatability |
0,01°Z |
light sensitivity |
Up to OD 5 |
Time |
About 4 seconds |
wavelength |
Standard configuration: 1 or 2 sets of wavelengths |
Measurement mode |
CONTINUE |
temperature
Measuring temperature |
Adapt the temperature sensor based on the rotation tube |
Sample temperature control |
External circulating water bath temperature control |
Temperature range |
Water bath temperature control: varies according to different circulating water baths |
Temperature Accuracy |
Different according to different circulating water baths/0.01 ° C |
Temperature accuracy |
Different according to different circulating water baths/± 0.03 ° C |
more parameters
polarimeter tube |
Up to 200mm in length |
ambient temperature |
+10 - +40 °C |
light source |
LED |
display |
7 "TFT touch screen, 800 x 480 pixels, 16 bit color |
operate |
Touch screen, keyboard *, mouse *, barcode scanner reader *, PC online* |
Interface/Online |
RS232 (1x), USB A (4x), USB B (1x), Ethernet (1x), analog, W-LAN/LAN* |
power consumption |
90 V / 50 Hz - 31,0 W |
standard model
model |
Saccharomat ® 103 T (587 nm) |