product details
simplebe situated between
SPMK518TintelligenceProcess calibratorIt is a desktop calibration instrument designed for high-precision and multifunctional calibration of thermal instruments, using high resolutionTFTColor screen, Chinese menu operation mode, powerful task management function, meet the requirements of calibrating temperature, pressure and other process instruments in industrial sites and laboratories, making it an ideal tool for calibrating, maintaining and repairing thermal instruments.
main features
Measurement and output are completely isolated and do not interfere with each other;
◆ It can measure voltage, current, resistance, frequency, pulse, switch (on/off), and can also use thermistor or thermocouple to measure temperature;
◆ Can output signals such as voltage, current, resistance, frequency, pulse, etc;
◆ Can simulate the measurement and output of 13 types of thermocouples and 11 types of thermal resistors, and provide cold junction compensation;
Provide a 24VDC power supply that is isolated from measurement and output;
Support data storage function, support step or programming output methods, etc;
◆built-inThe HART handheld device can manage basic information of HART type instruments, transfer ranges, calibrate zero and full range devices, and perform calibration and maintenance work;
◆ConnectableSDPModular intelligent digital pressure calibrator;
◆ Provide commonly used tools such as thermal engineering encyclopedia, analog transmitter, user built thermal resistance library, temperature and pressure unit conversion, etc;
◆ Automatic cold end compensation technology: embedded cold end insulation module, quickly tracks temperature changes;
High resolution TFT display screen, Chinese language display, good human-computer interaction interface;
RS232 interface and open communication protocol facilitate remote networking for users;
◆ Support online firmware upgrade.
Technical Parameter
Technical indicators for signal measurement(Environmental temperature 20 ℃± 5 ℃, one-year accuracy)
Signal type |
range |
accuracy |
resolution |
|
electric current |
(-30~30)mA |
±(0.01%RD+0.005%FS) |
0.1uA |
|
voltage |
(-30~30)V |
±(0.01%RD+0.005%FS) |
0.1mV |
|
Millivolt voltage |
(-75~75)mV |
±(0.01%RD+0.005%FS) |
0.1uV |
|
resistance |
3-wire system |
(0~400)Ω |
±(0.02%RD+0.005%FS) |
1mΩ |
4-wire system |
(0~400)Ω |
±(0.01%RD+0.005%FS) |
1mΩ |
|
frequency |
(1~50000)Hz |
±(0.005%RD+0.002%FS) |
0.1Hz |
|
pulse |
0~999999 |
±1A pulse |
1 |
|
Switch (on/off) |
If the switch is equipped with inspection voltage, the voltage range is(3~24)V |
|||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
|||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618) And custom types |
|||
pressure |
Refer to this websitePressure Range Selection Table |
Technical specifications for signal output(Environmental temperature 20 ℃± 5 ℃, one-year accuracy)
Signal type |
range |
accuracy |
resolution |
electric current |
0~22)mA |
±(0.02%RD+0.005%FS) |
0.1uA |
voltage |
0~12)V |
±(0.02%RD+0.005%FS) |
0.1mV |
Millivolt voltage |
(-10~100)mV |
±(0.02%RD+0.005%FS) |
0.1uV |
frequency |
0~50000.0)Hz |
±(0.05%RD+0.002%FS) |
0.1Hz |
pulse |
0~999999 |
±1individualpulse |
1 |
resistance |
0~400)Ω |
±(0.02%RD+0.005%FS) |
1mΩ |
(400~4000)Ω |
±(0.03%RD+0.01%FS) |
10mΩ |
|
24VDC |
(24±0.5)VMaximum load capacity50mA |
||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618) And custom types |
Other indicators
1. Operating environment:Environmental temperature (-10~50))℃; relative humidity<90% non condensing; Atmospheric pressure (86~106) KPa. a.
2. Storage environment:Environmental temperature(-20~60)℃; Storage humidity<90%Non condensing.
3. Power supply:220VAC.
4. External dimensions:(415×375×240)mm。
5. Weight:7.6kg.
Random accessories
packing list
name |
model |
unit |
quantity |
Intelligent process calibrator |
SPMk518T |
tower |
1 |
operation instructions |
《SPMK518TUser Manual |
stem or root of plants |
1 |
warranty card |
share |
1 |
|
certificate of conformity |
share |
1 |
|
power line |
strip |
1 |
|
Pen line |
0.5rice red |
strip |
1 |
Pen line |
0.5rice black |
strip |
1 |
Pen line |
1rice red |
strip |
3 |
Pen line |
1rice black |
strip |
3 |
Alligator Clip |
red |
individual |
3 |
Alligator Clip |
black |
individual |
3 |
packing list |
share |
1 |