product details
simple be situated between
The SPMK516 multifunctional temperature calibrator is a handheld thermal instrument calibrator that can measure and output electrical signals, simulate the measurement and output of thermocouple resistance, and more. Built in HART function, capable of calibrating and maintaining HART instruments. The mobile menu operation interface mode and powerful task management function make it an ideal tool for calibrating temperature, maintaining and repairing thermal instruments in industrial sites and laboratories.
main features
◆ It can measure voltage, current, resistance, frequency, pulse, switch (on/off), and can also use thermal resistors or thermocouples to measure temperature without interfering with each other;
◆ Can simulate13Thermocouples and11Measure and output thermal resistance, and provide cold end compensation;
◆ Provide isolation between measurement and output24VDCPower Supply;
◆ Support data storage function, support step or programming output methods;
◆ Provide commonly used tools such as thermal engineering encyclopedia, analog transmitter, user built thermal resistance library, temperature and pressure unit conversion, etc;
◆ Automatic cold end compensation technology: embedded cold end insulation module, quickly tracks temperature changes;
◆3.5High resolution display screen, Chinese language display, good human-computer interaction interface;
◆ Built inHARTHandheld device function, capable of achievingHARTCalibration and maintenance work for functional equipment such as basic information management of type instruments, range migration, zero and full calibration;
◆RS232The interface and open communication protocol facilitate remote networking for users and support online firmware upgrades.
Technical Parameter
Technical indicators for electrical signal measurement(Environmental temperature)20℃±5℃,One year accuracy).
Signal type |
range |
accuracy |
Resolution |
|
electric current |
(-30~30)mA |
±(0.01%RD+0.005%F.S) |
0.1uA |
|
voltage |
(-30~30)V |
±(0.01%RD+0.005%F.S) |
0.1mV |
|
Millivolt voltage |
(-75~75)mV |
±(0.01%RD+0.005%F.S) |
0.1uV |
|
resistance |
3Line system |
(0~400)Ω |
±(0.02%RD+0.005%F.S) |
1mΩ |
4Line system |
(0~400)Ω |
±(0.01%RD+0.005%F.S) |
1mΩ |
|
frequency |
(1~50000)Hz |
±(0.005%RD+0.002%F.S) |
1Hz |
|
pulse |
0~999999 |
±1A pulse |
1 |
|
Switch (on/off) |
If the switch is equipped with inspection voltage, the voltage range is(3~24)V |
|||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
|||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618)And custom types |
Technical specifications for signal output(Environmental temperature 20 ℃± 5 ℃, one-year accuracy)
Signal type |
range |
accuracy |
Resolution |
electric current |
0~22)mA |
±(0.02%RD+0.005%FS) |
0.1uA |
voltage |
0~12)V |
±(0.02%RD+0.005%FS) |
0.1mV |
Millivolt voltage |
(-10~100)mV |
±(0.02%RD+0.005%FS) |
0.1uV |
frequency |
0~50000.0)Hz |
±(0.005%RD+0.002%FS) |
0.1Hz |
pulse |
0~999999 |
±1A pulse |
1 |
resistance |
0~400)Ω |
±(0.02%RD+0.005%FS) |
1mΩ |
(400~4000)Ω |
±(0.03%RD+0.01%FS) |
10mΩ |
|
24VDC |
(24±0.5)V |
||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618) And custom types |
Other indicators
1. Operating environment: ambient temperature (-10~50) ℃; Relative humidity<90% non condensing; Atmospheric pressure (86~106) KPa. a.
2. Storage environment: ambient temperature (-20~60) ℃; Storage humidity<90% non condensing.
3. Power supply: AC/DC power supply.
Note: In order to achieve better measurement and output results for the calibrator, it is recommended to use battery power.
Random accessories
packing list
name |
model |
unit |
quantity |
Multi functional temperature calibrator |
SPMk516 |
tower |
1 |
operation instructions |
《SPMK516User Manual |
stem or root of plants |
1 |
warranty card |
share |
1 |
|
certificate of conformity |
share |
1 |
|
The power adapter |
individual |
1 |
|
Pen line |
0.5rice red |
strip |
1 |
Pen line |
0.5rice black |
strip |
1 |
Pen line |
1rice red |
strip |
3 |
Pen line |
1rice black |
strip |
3 |
Alligator Clip |
red |
individual |
3 |
Alligator Clip |
black |
individual |
3 |
Compensation wire |
Ktype |
strip |
1 |
Compensation wire |
Etype |
strip |
1 |
Compensation wire |
Utype |
cover |
1 |
packing list |
share |
1 |