The structural characteristics are:
1. Ozone generator system: mainly includes components such as ozone generator, air purification device, heat exchanger, and airflow control system. The ozone generator uses high-voltage silent discharge tubes, which can obtain higher concentrations of ozone. The sample aging chamber is enclosed and unlit (except for intermittent lighting).
2. A window is opened on the door of the test box to observe the surface of the sample. The outer shell of the test box is made of cold-rolled steel plate electrostatically sprayed, which has a beautiful and elegant appearance;
3. The inner walls of the chamber, conduits, sample fixtures, and structural materials are made of stainless steel that is not easily decomposed by ozone or corrosion-resistant inert materials.
4. Equipped with a timing device, the intelligent digital display controller is used for measuring and regulating ozone concentration. It can directly set the ozone concentration value and implement PID self-adjusting control, with high control accuracy and simple and convenient operation.
sketchmanufacturerThe conditions for use
1. Environmental temperature: 10-25 ℃;
2. Relative humidity: not greater than 85% R.H
3. Atmospheric pressure: 86-106Mpa
4. There is no strong vibration around;
5. No direct sunlight or direct radiation from other heat sources;
6. There is no strong airflow around. When the surrounding air is forced to flow, the airflow should not directly blow onto the box;
7. There is no strong electromagnetic field influence around;
8. There is no high concentration of dust or corrosive substances around.
9. To ensure the normal operation and convenience of the equipment, in addition to keeping the equipment horizontally placed, a certain amount of space should be reserved between the equipment and walls or objects.