Product Introduction:
According to the national regulations issued by relevant departmentsexecuteStandard, our company has successfully developed itGC-7900R Natural Gas Online Analyzer,comply withGB/T13610-2003Gas Chromatography for Composition Analysis of Natural GasSH/T0230-92The national analytical and testing implementation standards required by the "Liquefied Petroleum Gas Composition Determination Method (Chromatography Method)" and others. Widely used in gas companiesAluminum manufacturing, thermal power plants, ceramics, metallurgy, glass, papermakingWaiting for natural gas user units to analyze and test the calorific value of the natural gas used.
GC-7900R Natural Gas Online AnalyzerIt can be directly connected to natural gas pipelines to achieve online injection analysis, and the analysis results can be obtained in about 10 minutes. And it can achieve 24-hour continuous online monitoring, and the analyzed data results can be uploaded to the central control system through the user's network platform protocol. Providing users with accurate, fast, and concise detection methods, making the analysis and monitoring data clear at a glance.
GC-7900R Natural Gas Online Analyzer is suitable forThe component content analysis and detection of pipeline gases such as natural gas, liquefied gas, liquefied mixed air, and gas distribution. The instrument's built-in analysis software system can calculate characteristic indices such as high calorific value, low calorific value, density, relative density, Huabai number, and combustion potential of different gases in seconds.
GC-7900R Natural Gas Online AnalyzerFull analysis of oxygen, nitrogen, methane, ethane, propane, isobutane, n-butane, n-pentane, isopentane, and hydrocarbons with C6+or higher in natural gas can be performed in one injection.
Key Features:
1. Built in hydrogen generator, high electronic circuit integration,19 inch large screen industrial tablet computer display all-in-one machineGood reliability, simple operation,Ensure stable and reliable instrument quality, capable of long-term uninterrupted operation, which not only guarantees work efficiency but also saves maintenance costs. Automatically memorize operating parameters,Power on does not require resetting.
2. Equipped with a gas cut-off protection system to prevent instrument damage caused by misoperation.
3. The gas path system adopts a high-precision valve structure, which greatly improves the stability of the instrument gas flow rate, thereby obtaining well reproducible analysis results.
4. It can analyze the content of various components such as nitrogen, carbon dioxide, methane, ethane, propane, propylene, isobutane, n-butane, isobutene, n-butene, isopentane, n-pentane, etc. in natural gas and other gases. Content range: 0.01%~100%. Accurately calculate the calorific value, white number, density, etc. of gas by analyzing the content of each component.
5. It can be used in the laboratory or online.
6. The instrument has a unique fully automatic sampling system, which is connected to the outlet of the natural gas detection point to achieve 24-hour unmanned continuous sampling analysis and detection.
7Using the most advanced artificial intelligence software to operate the setting system,The detection data is uploaded to the user control room through the DCS system.
Technical Indicators:
1Temperature control indicators
column oven
Temperature control range: 6 ℃ to 400 ℃ at room temperature
Temperature control accuracy: within 200 ℃, the accuracy is ± 0.1 ℃
2Technical specifications of detector
Thermal conductivity pool detector(TCD)
Sensitivity: S ≥ 6000mV · ml/mg (n-hexadecane)
Noise: ≤ 20 μ V
Drift: ≤ 30 μ V/15 min
Linear range: ≥ 104
Configuration List:
GC-7900R Natural Gas Online Analyzer: Host+Thermal Conductivity Detector (TCD)+Tablet Computer (Built in)+Hydrogen Generator (Built in)
Data processing: dedicated chromatographic calorific value workstation (calorific value and component analysis)
Pipeline connection: Online natural gas pipeline connection device
Injector: Fully automatic injection valve
Data upload: 485, 212, DCS, etc
Chromatography column: 2 analytical columns
Sampler: 2L sampling bag
Startup tools: Work software startup package and usage tools