Application:
Used for online continuous monitoring of phosphate content in boiler water of steam power plants
Main features:
1. Using patented technology - a constant pressure controllable timing injection device, as a quantitative solution for water samples and reagents in the instrument; Abandon peristaltic pumps, capillaries, and air pressureQuantitative scheme for precision metering pumps. Greatly simplifying the quantitative system, reducing maintenance costs, and improving reliability.
2. Using patented technology, a composite photoelectric detection device has been developed, which includes a mixer, reactor, stirrer, colorimetric cup, and photoelectric colorimeter in chemical reactionsInternalization simplifies the structure and minimizes maintenance.
3. Periodic measurement and automatic cleaning methods eliminate bubble interference, accumulated pollution, sample crossover, and measurement lag that are difficult to solve with traditional technical solutionsStubborn illness greatly improves the stability of the instrument.
4. Use only one reagent.
5. Extremely low operating costs, with an annual cost of approximately 750 RMB per unit.
6. Complies with the DL/T677-2009 standard for the power industry of the People's Republic of China.
Working conditions:
1. Environmental conditions: Environmental temperature: 5~45 ℃ Relative humidity: ≤ 85% There are no corrosive gases in the surrounding air, no influential vibrations, and no strong electromagnetic interference.
2. Water sample conditions:
Water sample flow rate: 120-150ml/Min has no significant fluctuations, otherwise configure a constant current valve;
Water sample temperature: 15~45 ℃. There are no suspended solids in the water sample, otherwise a filter should be used.
3. Working power supply: AC220V ± 10%, 50Hz, 50W
If different power supplies are configured, they must be declared when ordering.
Technical indicators:
1. Measurement range: 0~1mg/L or 2mg/L (optional)
0~5mg/L,0~10mg/L,0~20mg/L
2. Basic error: ≤± 2% F · S
3. Stability: ≤± 1% F · S/24h
4. Reproducibility: Standard deviation
0-1mg/L or 2mg/L (optional) ≤± 0.02mg/L
0-5mgL≤±0.05mg/L
0-10mg/L≤±0.1mg/L,0-20mg/L≤±0.2mg/L
5. Silicon interference: 5mg/L SiO2Generate no more than 0.1mg/LPO43-positive interference
6. Reagent consumption: Monthly consumption of 2.8L
7. Analysis cycle: about 6 minutes or optional
8. Water sample channel: 1-6 channels (programmable)
9. Calibration method: automatic calibration
10. Output signal:
Current: 4-20mA (RL ≤ 500)Ω)
Alarm: normally open relay contact: 1A/24VDC or 1A/120VAC
Communication: Rs485
11Dimensions: Length x Width x Thickness: 640 × 440 × 315mm
Hole size: 612 × 412mm
12. Weight: 10kg
If there are special requirements, please specify them when placing an order