Guangdong Yuelian Instrument Co., Ltd
Home>Products>PCT-350 PCT High Voltage Accelerated Aging Test Chamber
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    13794910168
  • Address
    Building A, Tongzheng Smart Technology Park, No. 76 Tinghuan Road, Houjie Town, Dongguan City
Contact Now
PCT-350 PCT High Voltage Accelerated Aging Test Chamber
The PCT high-pressure accelerated aging test chamber is used to test the performance of products during storage, transportation, and use in high tempe
Product details

purpose

The PCT high-pressure accelerated aging test chamber is used to test the performance of products during storage, transportation, and use in high temperature, high temperature, high humidity, and pressure climate environments. It is mainly used to test the physical and other related properties of electrical and electronic products, components, parts, metal materials, and magnetic materials under simulated high temperature, high temperature, high humidity, and pressure climate conditions. After testing, the performance of the product can be judged through calibration to determine whether it can meet the requirements for product design, improvement, calibration, and factory inspection.

Box Structure

a. Outer wall material: stainless steel plate SUS304 #; Thickness: 1.2mm

b. Inner wall material: stainless steel plate SUS304 #; Thickness: 3.0mm

c. Box insulation material: fiberglass

d. Door insulation material: fiberglass

e. Circular inner box, stainless steel circular inner box structure, meets industrial safety standards and can prevent condensation and dripping during testing

f. Arc shaped inner liner, stainless steel arc shaped inner liner design, can avoid direct impact of steam overheating

g. Precise design, good air tightness, low water consumption, and can operate continuously for 999 hours with automatic water addition each time;

h. Automatic access control, anti pressure flat door with no external force automatic temperature and pressure safety detection, access control automatic locking control system, adopting advanced safety door handle design (door handle is safer without force), when the inner box has pressure greater than normal, the test door will be protected by anti pressure and cannot be opened; Protect the safety of operators.

i. Patent type packing, when the pressure inside the box is higher, the packing will have back pressure to make it more tightly integrated with the box body. Unlike the traditional compression type that relies on external force from outside to inside, when the pressure inside the box is 1Kg/cm2, the box door will be sucked inward with a suction force of 1Kg/cm2 to achieve a sealing effect; When the pressure inside the box is 2Kg/cm2, the box door will attract inward with a suction force of 2Kg/cm2 to achieve a sealing effect; The degree of compression of the box door is completely determined by the pressure inside the box So as to better protect the box door and extend its lifespan

j. Before the experiment begins, the equipment will automatically extract the unsaturated air from the original inner box, ensuring that each test sample is completed in a saturated vapor environment and ensuring the consistency of the test results

control system

a. Microcomputer saturated vapor LED digital temperature controller+time LED display controller, pressure pointer display meter

b. This system complies with the reliability test specifications for high-voltage accelerated aging: CNS, ISO, JIS, ASTM, DIN, BS, IEC, NACE, UL, MIL

c. Control object: Microcomputer+P.I.D.+S.S.R automatic calculation and control of saturated vapor temperature

d. Control mode: microcomputer control

Sample limitations

This experimental equipment prohibits:

Testing or storage of samples of flammable, explosive, and volatile substances;

Testing or storage of corrosive substance samples;

Testing or storage of biological samples;

Testing or storage of strong electromagnetic emission source samples;

Technical Parameter

product name PCT High Voltage Accelerated Aging Test Chamber
model PCT-35
Nominal internal volume 43L
inside dimensions 350 × 450 mm (φ × D) circular laboratory
External dimensions W900 XD800 XH1300mm
weight About 100kg
Test environment conditions Under the conditions of an ambient temperature of+25 ℃, relative humidity ≤ 85%, and no sample in the test chamber
temperature range 100 ℃~135 ℃ can be set arbitrarily
Humidity range 100% RH (saturated vapor humidity)
Pressure Range 1.5-3.0Kg/cm2 (including 1atm)
time frame 0 ~ 9999H
heating up time RT~120 ℃ for about 35 minutes
Rise time 0.0Kg/cm2~1.00Kg/cm2 within approximately 35 minutes
Temperature analysis accuracy Temperature: 0.1 ℃
temperature control accuracy Temperature: ± 0.5 ℃
Pressure control accuracy ±0.2Kg/cm2
temperature uniformity Temperature: ± 2%
Circulation mode Natural convection cycle of water vapor
Box door Anti pressure flat door without external force
control panel Temperature controller display screen, time setter, operation switch, pressure gauge, alarm indicator light
Distribution control cabinet Distribution board, exhaust fan
heater Titanium alloy heating tube
Power cord hole Located on the back of the box
temperature controller Imported LED display precision temperature controller
Screen display function Direct display of temperature setting (SV) and actual (PV) values
Chamber Adjustable over temperature and low water level protection; Adjust the channel limit to exceed the temperature limit; Overpressure protection; Leakage protection; Overload and short circuit protection
Installation site The ground is flat and well ventilated; There is no strong vibration around the equipment; There are no flammable, explosive, corrosive substances or dust around the equipment; Leave appropriate space for use and maintenance around the equipment
environmental condition Temperature: 5 ℃~30 ℃; Relative humidity: ≤ 85%; Air pressure: 86kPa~106kPa
power supply AC220V 1-phase 3-wire
Voltage allowable fluctuation range: ± 5%
Allowable frequency fluctuation range: 50 Hz ± 0.5Hz
The grounding resistance of the protective grounding wire is less than 4 Ω
Require users to independently configure air switches of corresponding capacity for the equipment at the installation site
Installed power About 2.5 Kw
Maximum current 10 A
Air Switch 32 A

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!

Successful operation!

Successful operation!