Oxi3310 handheld dissolved oxygen analyzer (WTW, Germany)Product Introduction:
1. Capable of measuring dissolved oxygen, saturation, and temperature, with a maximum resolution of 0.01mg/l;
2. Interval control recording function, can accommodate up to 5000 records, can set storage intervals between 1s-60min, manually store 200 sets of measurement data, each set of data includes measurement values, temperature, date, time, number, and calibration status during measurement, etc;
3. The concise text menu ensures comfortable and convenient operation;
4. The instrument casing is extremely sturdy, and the silicone keypad is 100% waterproof and easy to clean;
5. Touch feedback buttons ensure accurate and error free operation;
6. Built in calibration timer to ensure stable and reliable data results;
7. IP66/IP67 is suitable for outdoor use under any weather conditions;
8. The calibration principle using saturated humid air method requires very little maintenance;;
9. After calibration, the electrode status will be automatically displayed on the screen;
10. Safe testing - ATC automatic temperature compensation;
11. The battery life is up to 1000 hours, and it lasts for 100 hours when the background light is constantly on;
12. The new Continuous Test Control (CMC) function ensures the reliability of all readings: 1) When the measured value is within the calibration range, an indication will be given on the display; 2) When exceeded, the display will sound an alarm;
13. 100% waterproof USB interface, enabling fast data transfer between personal computers and handheld computers, suitable for outdoor use;
14. Excellent technical specifications, with a maximum range of 90mg/l;
Product Configuration
Includes Oxi3310 host, portable case, cleaning solution, electrolyte, replaceable film head, grinding sheet, bracket, beaker, operation manual, and battery, etc.
Oxi3310 handheld dissolved oxygen analyzer (WTW, Germany)Technical parameters:
model |
Oxi 3310 |
D. O. Concentration |
0.00-19.99 mg/l ± 0.5% test value 0-90 mg/L ± 0.5% test value |
saturation |
0.0... 199.9% ± 0.5% test value 0-600% ± 0.5 test value |
partial pressure |
0-199.9hPa, 0-1250hPa ± 0,5% |
temperature |
-5.0-105.0°C ± 0.1°C |
Automatic reading |
automatic/Manual Selection |
correct |
CellOx/DurOxProbe,OxiCalCalibration kit |
Correction record |
Record the last 5 corrections |
Pressure Compensation |
Automatic compensation, 500-1100 mbar |
display |
7-segment LCD chart display with backlight |
data storage |
hand movement200/automatic5000 |
Data recording method |
hand movement/time driven |
interface |
USB interface |
power supply |
4 x 1.5V alkaline battery Or four 1.2V rechargeable batteries |
Continuous operation time |
Up to 1000 hours (150 hours with backlight) |