VIP member
Ordinary pipette
Model inspection point (ul) Capacity allowable error (%) Repeatability error (%) QY-2 (0
Product details
model |
Detection point (ul) |
Capacity allowable error (%) |
Repetitive error (%) |
|
|
QY—2 (0.1---2ul) |
0.2 |
±12.00 |
≤6.00 |
|
1 |
±4.00 |
≤2.50 |
||
2 |
±4.00 |
≤1.50 |
||
|
QY—10 (0.5---10ul) |
1 |
±4.00 |
≤2.50 |
|
5 |
±2.00 |
≤1.00 |
||
10 |
±1.50 |
≤0.80 |
||
|
QY—20 (2---20ul) |
2 |
±5.00 |
≤1.50 |
|
10 |
±2.00 |
≤1.00 |
||
20 |
±2.00 |
≤0.80 |
||
|
QY—50 (5---50ul) |
5 |
±2.50 |
≤1.50 |
|
25 |
±1.00 |
≤0.50 |
||
50 |
±1.00 |
≤0.40 |
||
|
QY—100 (10---100ul) |
10 |
±2.00 |
≤1.00 |
|
50 |
±1.20 |
≤0.50 |
||
100 |
±1.20 |
≤0.30 |
||
|
QY—200 (20---200ul) |
20 |
±2.00 |
≤0.80 |
|
100 |
±1.00 |
≤0.70 |
||
200 |
±1.00 |
≤0.50 |
||
|
QY—1000 (100---1000ul) |
100 |
±1.50 |
≤0.80 |
|
500 |
±0.90 |
≤0.40 |
||
1000 |
±0.90 |
≤0.28 |
||
|
QY—5000 (500---5000ul) |
500 |
±1.00 |
≤0.50 |
|
2500 |
±0.50 |
≤0.20 |
||
5000 |
±0.60 |
≤0.20 |
||
|
QY—10000 (1000---10000ul) |
1000 |
±1.00 |
≤0.50 |
|
5000 |
±0.60 |
≤0.20 |
||
10000 |
±0.60 |
≤0.20 |
||
Online inquiry