The TR 6900 multi parameter water quality analyzer is our latest designed specialized detection instrument for enterprise pollution source wastewater, which can quickly measure parameters such as COD, ammonia nitrogen, total phosphorus, heavy metals, turbidity, SS, etc. in enterprise wastewater. The instrument interface design is simple and elegant, with a fixed multi-channel optical path design mode that can automatically switch light source channels, making the light source more stable and eliminating the error factors of traditional manual rotation.
This instrument has strong scalability and provides convenience for later enterprises to increase detection parameters. This instrument is developed and designed based on authoritative certification methods, and the detection principles of each parameter comply with the recognized methods in the environmental protection industry. The instrument is characterized by its fast detection, simple operation, and low cost. Widely used in emergency monitoring, sewage treatment, chemical, pharmaceutical, hospital wastewater, food, printing and dyeing industries for wastewater detection, and can also be applied to wastewater research and use in scientific research institutions, colleges and universities, and other institutions.
Measurement parameters | COD | Ammonia nitrogen | total phosphorus | total nitrogen | turbidity |
Measurement range | 0-10000mg/L | 0-50mg/L | 0-20mg/L | 0-100mg/L | 0-1000NTU |
Anti chlorine interference | 1000mg/L | - | - | - | - |
limit of detection | 5mg/L | 0.01mg/L | 0.02mg/L | 0.5mg/L | 0.5NTU |
Dissolve the environment | 165℃,15min | No need to dissolve | 125℃,30min | 125℃,30min | No need to dissolve |
Measurement Error | ≤± 5% | ||||
repeatability | ≤±5% | ||||
Optical stability | ≤± 0.001A/20 minutes (100000 hours lifespan) | ||||
Colorimetric method | Colorimetric tube (digestion colorimetric integrated tube) | ||||
Batch processing volume | 16, 25 (optional) | ||||
data processing | 30000 detection records, 96 curves (supporting user-defined calibration) | ||||
transmission mode | USB data cable, LAN, etc | ||||
printer | Built in thermal printer | ||||
Operation interface | Chinese/English (optional) | ||||
display | 5-inch color LCD screen | ||||
Power supply mode | AC(220V±10%),50Hz | ||||
power | 5W | ||||
Host size | 348mm*250mm*145mm | ||||
Instrument weight | Host: 2.6kg; Dissolver: 6kg | ||||
ambient temperature | 5~40℃ | ||||
ambient humidity | ≤ 85% non condensing |
Users choose parameters according to their own business needs
Serial Number | Measuring Item | measuring range | limit of detection |
1 | COD (high range) | 0-10000 mg/L | 5 |
2 | COD (low range) | 0-200 mg/L | 5 |
3 | Manganese based COD | 0-10 mg/L | 0.1 |
4 | Ammonia nitrogen | 0-50 mg/L | 0.05 |
5 | total phosphorus | 0-20 mg/L | 0.02 |
6 | total nitrogen | 0-100 mg/L | 0.5 |
7 | turbidity | 0-1000NTU | 0.5 |
8 | Suspended solids (SS) | 0-500 mg/L | 5 |
9 | chroma | 0-500 degrees | 5 |
10 | hexavalent chromium | 0-5 mg/L | 0.01 |
11 | total chromium | 0-5 mg/L | 0.01 |
12 | copper | 0-50 mg/L | 0.05 |
13 | nickel | 0-4 mg/L | 0.1 |
14 | zinc | 0-15 mg/L | 0.05 |
15 | iron | 0-50 mg/L | 0.01 |
16 | phosphate | 0-20 mg/L | 0.02 |
17 | sulfate | 0-200 mg/L | 5 |
18 | sulfide | 0-4mg/L | 0.01 |
19 | cyanide | 0-1 mg/L | 0.004 |
20 | fluoride | 0-10 mg/L | 0.02 |
21 | nitrate nitrogen | 0-10 mg/L | 0.5 |
22 | nitrite | 0-1 mg/L | 0.002 |
23 | Residual chlorine | 0-10 mg/L | 0.1 |
24 | total chlorine | 0-10 mg/L | 0.1 |
25 | chlorine dioxide | 0-10 mg/L | 0.05 |
26 | silica | 0-5 mg/L | 0.1 |
27 | Volatile phenol | 10 mg/L | 0.1 |
…… | …… | …… | …… |
Performance characteristics
1. Display interface: Large LCD screen display, button operation, direct data reading, simple and time-saving operation.
2. Wide detection range: dozens of detection parameters can be expanded and customized according to needs.
3. Light source advantage: using imported cold light source, optical performancegoodNo preheating required, with a lifespan of up to 100000 hours.
4. Intelligent operation: Intelligent operation program that guides users to easily complete operations.
5. Diversified testing methods: Standard measurement or continuous measurement mode can be selected, freely choosing between speed and accuracy.
6. Operation safety: Using a digestion colorimetric integrated tube, the measurement is simple, fast, and safe.
7. Quick printing: Built in printer that can print current data and historical measurement data.
8. Data processing: Memory standard working curve and 96 user-defined curves, capable of storing 30000 pieces of detection data.
9. Professional testing: Equipped with specialized testing reagents to minimize user operation and make testing more accurate.
10. Low usage cost: low consumable prices and minimal usage.
Online operation function
Supports network port and USB interface, capable of data transfer with PC end.
Software advantages
1. Supports Chinese/English display (optional), designed to be friendly, intelligent, and easy to operate.
2. Built in system program and user program, which can add frequently tested items to the user program for easy searching.
3. It has data storage and power-off protection functions to prevent data loss and facilitate the retrieval of historical measurement data.
4. Built in standard operating procedures, no need for user calibration and calibration.
5. Users can also develop their own user programs and reagents to adapt to different situations and water conditions.
6. Built in help interface, explaining the general precautions for user detection process.
7. Equipped with a one click factory reset function.
Advantages of digestion instrument
1. Standard 16 hole digestion analyzer, optional 25 hole digestion analyzer,moreTo meet the needs of users to a large extent.
2. Adopting intelligent PID temperature control technology and dual anti overheating protection system, heating is safe, uniform, and fast.
3. Built in standard digestion program, one click digestion design, only need to press the confirm button to complete the entire digestion process.
4. Support custom modes, allowing for free setting of digestion temperature and time, achieving one click digestion in the target mode.
5. Used for the digestion of COD, total phosphorus, total nitrogen, total chromium and other projects.