Product Introduction:
The multifunctional hydrogen rich machine can produce high-purity hydrogen gas, with a purity of up to 99.999%. By absorbing hydrogen gas, a large number of hydrogen molecules can enter the body and exert anti-cancer effects. A hydrogen molecule is like a warrior, eliminating free radicals in the body one by one and surrounding cancer cells. With these warriors, cancer cells will become anxious and fearful. Because hydrogen molecules have extremely strong penetration ability and can quickly reach every corner of the body, cancer cells will eventually have nowhere to hide in the body and disappear into thin air.
1RX-FQJ300/600Main features of export type multifunctional hydrogen enrichment machine
International Advanced“SPE”(Solid state electrolysis)technology)
Alkali free, direct electrolysis of pure water
DuPont Ionic Membrane
Electrolytic cell: made of titanium alloy material with iridium tantalum coating
Output flow displayed on LCD screen
IIRX-FQJ300/600Type technical parameters
1Hydrogen purity:99.999%
2Hydrogen flow rate: 0-300ml/min0-600ml/min
3Output pressure:0.2MPa
4Available pure water: greater than2MΩDeionized water or secondary distilled water
5Pressure stability:<0.001Mpa
6Power supply:220V±10% 50Hz
7Power consumption:150W/200W
8External dimensions:400mmX270mmX430mm
9Net weight: 10kg /12kg
3、 Product Features:
1Can generate99.999%Hydrogen gas has no corrosion to the instrument
2The electrolytic cell adopts imported ion membrane, which has drug resistance, durability, and long service life
3The output flow of the electrical automatic control system is stable with minimal fluctuations.
4The voltage of the electrolytic cell is low (reaching:2.20-2.30).
5The machine is equipped with an anti water circuit design inside.
Instrument usage method
1. Open the bucket lid and add deionized water or distilled water (Wahaha purified water bottle)The water storage capacity of Yibao purified water should be maintained atBetween 1600ml and 2000ml.
2. Before starting up, please pull out the drain pipe and release a small amount of waterUsed to exhaust the air inside the pipe and ensure smooth water flow in the electrolytic cell.
3. Screen operation instructions: "+" indicates shutdown, "-" indicates startup
(1) Start button: Under normal conditions, press the "-" button to start producing hydrogen gas, while the "second" symbol flashes. Press the "+" key to turn off the hydrogen generator, while the "second" symbol does not flash.
(2) Set traffic and time:
pressPress the "Settings" button, the hydrogen flow rate will flash, and start adjusting the gas flow rate. press + The key value increases, pressThe value of the "-" key decreases.
Press the "Settings" button again, and the timer "hour" will flash to start adjusting the hours. Press the "+" button to increase the value, and press the "-" button to decrease the value.
Press the "Settings" button again, and the timer "minute" will flash, starting to adjust the minutes. Press the "+" button to increase the value, and press the "-" button to decrease the value.
Press the "Settings" button again, save the settings, restore normal operation, the second symbol flashes, and hydrogen production begins.
4Turn on the instrument switch and the instrument will start working normally to produce gas. Insert the randomly equipped nasal cannula into the hydrogen outlet on the front panel to breathe hydrogen gas or produce hydrogen rich water.
5Cut off the forked part of the randomly brought nasal cannula with scissors and replace it with a hydrogen conducting tube with one end in and one end out. Put the hydrogen tube into the water cup and let the hydrogen gas gurgle in the water5-10 minutes of gas will produce a content of800 to1200Hydrogen rich water at around ppb.
6The water tank should be kept clean, 3 monthsClean the water tank once on both sides and add new purified water.
4、 Instrument precautions
1It is strictly prohibited to use in areas with open flames nearby
2Good ventilation conditions
3It is strictly prohibited to use water outside the regulations for instrument operation, such as tap water and mineral water.
4Do not dismantle the electrolytic cell at will
5、 Daily maintenance and troubleshooting of instruments
Fault phenomenon |
Reason for malfunction |
Inspection and exclusion methods
|
The instrument cannot work when turned on The power light is not on |
1. The circuit is not connected 2. The fuse is burnt out |
Connect the circuit properly Replace the fuse |
The pressure of hydrogen production cannot be increased The flow display is normal, and there are no bubbles generated when the hydrogen absorption tube is placed in water |
Air leakage in the internal pneumatic system of the instrument
|
Use detection solution to check the connections of each gas path Tighten the leakage point |
Output flow display 000, gas production light on |
Poor quality of working water leads to aging of the electrolytic cell Low water level |
Replace with new working water and clean the water tank or replace the electrolytic cell Raise the water level |
Numerical value of hydrogen production flow rate The deviation is too large |
1. Aging of electrolytic cell 2. Abnormal power output current |
1. Replace and clean the electrolytic cell 2. Adjust power supply, replace power supply |
There is water in the purification and drying tube |
The soda water separator cannot drain properly |
Replace the air-water separator Clean the air-water separator |
Excessive consumption of pure water |
1. There is leakage in the water circulation pipeline 2. The temperature of the electrolytic cell is too high |
Observe and eliminate water leakage Replace or clean the aging electrolytic cell |