nProduct specificationsparameter
focal length |
TR 555:3×500mm TR 557:2×500mm+750mm TR 777:3×750mm |
optical system |
Addition and subtraction dual C-T Aberration correction system with high-throughput multi light source system |
detector |
CCD/EMCCD/InGaAs/PMT semiconductor refrigeration camera and liquid nitrogen refrigeration camera are optional |
lowWavenumber range |
<±10cm-1Use a low bandwidth filter |
High wave number range |
9000cm-1(@532nm) |
Stray light suppression |
10-9 Or even better |
microscope |
Zhengli Olympus microscope BX43 BX53、BX51WI, Inverted Olympus microscope IX73 and dual channel microscope |
objective |
Long or short focal length UV, VIS,NIR |
Resolution (FWHM) |
<0.2cm-1@633nm |
spatial resolution |
<1 μ m horizontally< 2 μ m longitudinal |
Low frequency range |
<±10 cm-1Use a low bandwidth filter |
Step size of piezoelectric/mechanical XYZ platform |
<10nm /<50nm |
temperature |
Refrigeration -196 ° C -600 ° C, heating limit 1500 ° C |
Ultra narrow bandwidth filter |
488, 514, 532, 633, 785, and 1064nm |