LCZD-1008 multifunctional fully automatic oscillating instrument, in accordance with the oscillation degassing method of GB/T17625-1998 "Gas Chromatography Determination of Dissolved Gas Components in Insulating Oil"; The oscillation heating method in accordance with DL429.4-91 "Quantitative Determination of Water Soluble Acids" is used. The instrument is controlled by a microcomputer program and is used for constant temperature and timed heating, oscillation, and degassing of various liquids in the laboratory.
Instrument characteristics
240 × 128 dot matrix LCD display screen, no labeled button operation;
Adopting microcomputer control, PID self-tuning temperature control method, temperature control * *;
Mechanical transmission is anti fatigue, anti vibration, maintenance free, and low-noise;
Equipped with an external thermometer socket for easy calibration of temperature by the operator;
Small size, light weight, and exquisite appearance;
Can hold 8 100ml syringes or 4 250ml triangular bottles simultaneously;
Can achieve full Chinese/full English interface display
Technical Parameter
Temperature range: Room temperature to 120 ℃;
Temperature control accuracy: room temperature to 50 ℃± 0.2 ℃;
50℃~100℃±0.3℃
100℃~120℃±0.5℃
Oscillation frequency: 275 ± 3 times/minute
Oscillation amplitude: 35mm
Chromatographic oscillation degassing: oscillate for 20 minutes, stand still for 10 minutes;
Constant temperature of 50 ℃
Water soluble acid oscillation: oscillate for 5 minutes at a constant temperature of 75 ℃;
Other physical oscillations: any set oscillation within 0-99 minutes
Static time ± 5 seconds;
Set any temperature within the range of room temperature to 120 ℃
Noise:<35 decibels
Power supply voltage: AC220V ± 20% 50HZ ± 10%
Applicable ambient temperature: 5 ℃~50 ℃
Applicable environmental humidity: ≤ 85% RH
Dimensions: 500 × 350 × 385 (mm);
Weight: Approximately 35Kg