intelligent typeTracking Test ApparatusLDQ-2C
Product Overview:
The leakage trace tester is based onIEC60112-2003 GB4706.1-2008 GB4207-2003GB/T6553-2003 GB7000.1-2003Manufacturing based on standard design.
Technical Specifications:
1. Control method:PLCtouch screen;
2. Electrode material: Copper+platinum(platinum12mmlong);
3. Each electrode exerts a force on the sample:1.0±0.05N;
4. Droplet time:30s±0.1s;
5. Apply voltage:100~600V(48~60Hz)Adjustable between, short-circuit current is within1.0±0.1AThe voltage drop does not exceed10%;
6. Dropping device: No adjustment is required during the experiment;
7. Drop height:30~40mm;
8. Drop size:44~55drop/1cm³
9. Control droplets using solenoid valves,Accurate control of droplet size, twentyThe maximum drop accuracy can be achieved0.02g;
10. Circuit judgment: Short circuit current greater than0.5ATime maintenance2The relay operates in seconds, cutting off the current and indicating that the test sample is not qualified;
11. Internal volume:0.5m3(Optional)0.75m3or1m3);
12. External dimensions:wide1170mm*deep630mm*high1330mm;
13. Box material: Iron plate spray coating;
14. Exhaust hole:Ø100mm;
15. Working power supply:AC 220V 50HZ 5A