Performance characteristics
·Instrument level: 0.5 level, conductivity measurement range is displayed in 6 segments;
·Automatic calibration from one to four o'clock, with calibration guidance function;
·Preset calibration time and remind
·Self diagnostic function, standard solution exceeding the standard and pollution alarm prompt; And provide corresponding information; ★
·Can use a total of 8 standard solutions from CH/USA for calibration and custom standard solutions for calibration;
·Electrode constant selection; Temperature compensation coefficient setting; ★
·Ultra pure water (ultra-low conductivity) stable measurement mode;
·Equipped with stable endpoint display and automatic locking display mode, optimized reading accuracy and response time;
·90 × 68mm large screen LCD, large font display, white backlight, real-time clock and date display, clear and easy to operate;
·The operation interface is simple; Touch sensitive panel buttons, moisture-proof, dust-proof, and oil proof, with a long service life; ★
·(0-50.0) ℃ automatic/manual temperature compensation;
·500 sets of data storage; ★
·Equipped with USB communication interface; Standard data CD; After data transmission to the computer, the measurement serial number, date, time, measurement value, and temperature value can be displayed; ★
·Configure conductive electrodes with stable performance and temperature compensation;
·360 degree steering and vertical lifting universal electrode bracket;
·The temperature unit can be selected as ℃ or ℉;
·The instrument can be reset to factory settings with one click ★;
·Compliant with IP54 dust and splash protection level, the instrument socket is sealed with a rubber cap for protection.
measuring range |
<20 μS/cm |
0.5μS/cm~100mS/cm |
>100mS/cm |
||
|
conductance electrodeconstant |
K=0.1 |
K=1.0 |
K=10 |
||
Calibration solution |
146.6μS/cm |
146.6 μS/cm |
1408 μS/cm |
12.85 mS/cm |
111.3 mS/cm |
Electrode model |
E3224Tconductanceelectrode |
E2214TandE2224Tconductanceelectrode |
E1214Tconductanceelectrode |
||
Explanation: Testing conductivity<1.0 μS/cmHigh purity water should be subjected to flow testing using a flow cell.
Technical Parameter
instrumentimplementlevel |
1.0level |
|
measureQuantity rangeenclose |
conductancerate |
0~200.0mS/cmIncluding five sections: (0.00~20.00 μ S/cm) (20~200.0μS/cm)(200~2000μS/cm) (2.00~20.00mS/cm)(20.0~200.0mS/cm) |
electricityResistivity |
(0~100)MΩ•cm |
|
TDS |
(0~100)g/L |
|
saltlinear measure |
(0~100)ppt |
|
warmlinear measure |
(-10~110)℃ |
|
resolution |
conductancerate |
0.01/0.1/1μS/cm; 0.01/0.1mS/cm |
warmlinear measure |
0.1℃ |
|
basicmistakediffer from |
conductancerate |
±1.0%FS |
warmlinear measure |
±0.5℃ |
|
Standard configuration
CD510Standard host,E3224TGlass conductive electrode(K=0.1), circulation pool,Buffer solution set, universal electrode holder,USBCommunication cable, communication software CD,9VThe power adapter