Product features:
LHH-250GP Drug Strong Light Stability Test Chamber Baidian Instrument
◆ Humanized design
The brand new fluorine free design ensures high efficiency, low energy consumption, and promotes energy conservation, keeping you at the forefront of a healthy lifestyle.
● Microcomputer controller, stable, accurate and reliable control, using 304 stainless steel inner liner, four corner semi-circular arc, easy to clean and operate.
Unique air duct circulation ensures even distribution of wind inside the studio. There is a 25mm diameter test hole on the left side of the box.
Product features:
LHH-250GP Drug Strong Light Stability Test Chamber Baidian Instrument
The brand new fluorine free design ensures high efficiency, low energy consumption, and promotes energy conservation, keeping you at the forefront of a healthy lifestyle.
● Microcomputer controller, stable, accurate and reliable control, using 304 stainless steel inner liner, four corner semi-circular arc, easy to clean and operate.
Unique air duct circulation ensures even distribution of wind inside the studio. There is a 25mm diameter test hole on the left side of the box.
◆Continuous operation guarantee
Two sets of imported compressors automatically switch to ensure that drug testing runs continuously for a long time without any malfunctions. Breakthrough the defect of domestic drug testing chambers being unable to operate continuously for a long time.
Continuous operation does not require defrosting, avoiding temperature and humidity fluctuations inside the box caused by defrosting during use.
◆ Quality assurance
Key components such as temperature and humidity controllers, compressors, and circulating fans are all imported products, which have the characteristics of long-term stable operation, safety, and reliability.
◆ Safety features
● Independent temperature limit alarm system, capable of sound and light alarm to alert the operator, ensuring safe operation of the laboratory without accidents.
● Alarm for low or high temperature and over temperature, alarm for high or low humidity.
● It has a password lock screen function to avoid accidental operation by non experimental personnel.
◆ Imported humidity sensor
Choose a humidity sensor that can operate at high temperatures to avoid the hassle of frequent replacement of wet and dry bulb bands.
UV sterilization system (optional)
The ultraviolet sterilization lamp is placed on the back wall of the box, which can regularly disinfect the inside of the box and effectively kill floating bacteria in the circulating air and humidification tray water vapor inside the box, thereby effectively preventing contamination during drug testing.
◆ Automatic monitoring and control of illuminance (optional)
Breaking through the defect of the inability to monitor and control the illumination of existing domestic stable test chambers, using light sensors for monitoring and stepless adjustment, reducing the attenuation of illumination and test errors caused by lamp aging.
◆ Data recording and fault diagnosis display
When the test box malfunctions, the dynamic display screen will show fault information, and the operational faults of the test box will be clear at a glance.
It can be connected to a printer or 485 communication interface to record temperature and time curves using a computer and printer, providing strong guarantees for data storage and playback during the experimental process.
Programmable touch screen controller (optional)
Adopting a large touch screen display, the screen operation is simple and the program editing is easy.
The controller operation interface is available in both Chinese and English, and the real-time operation curve chart can be displayed on the screen.
● It has a capacity of 100 sets of programs, 1000 segments, and 999 loop steps, with a maximum set time of 99 hours and 59 minutes per segment.
After inputting the data and experimental conditions, the controller has a screen lock function to avoid manual touch and shutdown.
Equipped with P.I.D automatic calculation function, it can immediately correct temperature and humidity change conditions, making temperature and humidity control more efficientstable.
Equipped with RS-232 or RS-485 communication interface, it can be programmed on a computer to monitor the testing process and perform functions such as power on/off.
name |
Drug Stability Test Chamber |
Comprehensive drug stability test box |
Drug Strong Light Stability Test Chamber |
||||
model |
LHH-80SD LHH-150SD LHH-250SD |
LHH-80SDP LHH-150SDP LHH-250SDP |
LHH-150GSD LHH-150GSP LHH-250GSD LHH-250GSP |
LHH-150GP LHH-250GP LHH-400GP |
|||
Temperature Control Range |
0~65℃ |
No light0~65℃ Illuminated10~50℃ |
|||||
temperature fluctuation/Uniformity |
±0.5℃/±2℃ |
||||||
Humidity range/deviation |
35~95%RH/±3%RH |
not have |
|||||
Light intensity/error |
not have |
0~6000LXAdjustable ≤±500LX(Infinite dimming) |
|||||
time range |
each paragraph1~99hour |
||||||
Temperature and humidity regulation method |
BTHC |
Balanced temperature regulation method |
|||||
refrigeration system/Cooling method |
Two sets of independent original imported fully enclosed compressors with automatic switching(LHH-80SD/80SDPFor a set of independent original fully enclosed compressors) |
||||||
controller |
Programmable LCD Controller |
Programmable touch screen controller |
GSDProgrammable LCD Controller GSPProgrammable touch screen controller |
Programmable LCD Controller |
|||
sensor |
Pt100platinum Capacitive humidity sensor |
Pt100platinum |
|||||
Ambient Temperature |
RT+5~30℃ |
||||||
power supply |
AC220V±10% 50HZ |
||||||
power |
2000W/2100W/2300W |
2250W/2250W/2500W/2500W |
1450W/1700W/3200W |
||||
volume |
80L 150L 250L |
80L 150L 250L |
150L 250L |
150L 250L 400L |
|||
|
Inner container size(mm) W×D×H |
400×400×500 550×405×670 600×500×830 |
400×400×500 550×405×670 600×500×830 |
550×405×670 600×500×830 |
550×405×670 600×500×830 700×550×1140 |
|||
|
External dimensions(mm) W×D×H |
550×790×1080 690×805×1530 740×890×1680 |
550×790×1080 690×805×1530 740×890×1680 |
690×805×1530 740×890×1680 |
690×805×1530 740×890×1680 950×850×1850 |
|||
Load tray(standard configuration) |
2/3/3piece |
2/3/3piece |
3/3piece |
3/3/4piece |
|||
Embedded printer |
standard configuration |
standard configuration |
standard configuration |
standard configuration |
|||
Safety device |
Compressor overheat protection, fan overheat protection, over temperature protection, compressor overpressure protection, overload protection, water shortage protection. |
||||||
notes |
1、SD、SDP、GSD、GSP、GPSeries, standard embedded printer 2Printer with touch screen controller485Interface andUSBChoose two out of three interfaces,LHH-80SD、LHH-150SD/GSD,LHH-250SD/GSDPrinter and485Choose one of the two interfaces. 3、GP、GSD、GSPSeries, manual infinite dimming, standard illuminance monitor, built-in top light emitter 4、GP、GSD、GSPSeries, optional two-layer light emitter 5、LHH-80SD、LHH-80SDPFor a set of independent original fully enclosed imported compressors |
||||||
|
|||||||
|
|
|
|
|
|
|
|