KHR3000Digital convex nose structure Rockwell hardness testing machine
Experimental process:● Loading, holding, and unloading are precisely controlled by the computer
Compatibility standards:● Latest national standard:GB/T230.2、● Latest international standards:ISO6508,BSEN6508
Domestic pioneering:● The first domestically developed aluminum and aluminum alloy with copper copper alloy conversion function.● The test force value automatically switches without the need for manual switching.
Production process:Adopting high-precision sensors imported from Germany, with measurement resolution0.1HR
Structural features:● Featuring a unique cantilever pressure head structure,commonly known as"Protruding nose"structure,Can test the surface of workpieces that cannot be directly measured by ordinary hardness testers,Such as the inner surface, concave convex surface, groove surface, etc. of the workpiece,No need to damage the sample,It can complete the testing.
Auxiliary functions:Data processing: automatic mean, mean square error, maximum and minimum values; ● Free conversion of hardness standards;
● Tolerance setting curvature correction; ● Dual display in both Chinese and English; Applicable: Hard alloy, carbon steel, alloy steel, steel and cast steel, alloy tool steel, stainless steel, gray cast iron, ductile iron, cast aluminum alloy, copper Zinc alloy (brass), copper tin alloy (bronze), pure copper, forged steel, engineering plastics, etc.
Technical Specifications:
model |
KHR3000 |
Hardness scale |
rockwellA,B,C,D,E,F,G,H,K, L,M,P,R,S,V,together15Type ruler |
resolution |
0.1HR |
Initial testing force |
98.07N/10kgf |
Total test force |
588.4N/60kgf,980.7N |
display |
Large screen, high definition, high brightness, backlit dot matrixLCD |
menu text |
Chinese-English menu |
Operation Method |
Membrane keypad |
test process |
autocomplete |
Test force loading time |
2-50Seconds, can be set and dynamically displayed, stored |
Accessibility |
Upper and lower limit settings, over tolerance discrimination alarm; Data statistics, calculate the average, standard deviation, maximum, and minimum values; Surface correction, automatic correction of cylindrical and spherical measurement results |
Scale conversion |
The test results can be converted intoHBSTheHBWTheHVTheHLDTheHKTheHR15NTheHR30NTheHR45NTheHB10TheHB30TheHVTheHLDThe σbTheHKTheHR15TTheHR30TTheHR45TTheHB(Cu)TheHV(Cu)TheHB(Al)TheHV(Al)and σb(Al)TheHB(Al-d)and σb(Al-d) |
data storage |
Powerful data storage function: automatic storage of test time, test serial number, and test results, which can be called, viewed, and printed at any time, Cannot be changed or deleted. Effectively monitor testing history to ensure the objectivity and fairness of historical data |
execution standard |
GB/T230.2,ISO6508,BSEN6508,ASTM E-18 |
Maximum testing space |
vertical direction260mmIn the horizontal direction150mm |
Types of test parts |
Plane; Cylindrical surface, minimum outer diameter3mm; Inner ring surface,minimum inner diameter23mm |
storage |
auto-storage500Group measurement data, time, and test sequence number |
communication interface |
USBTheRS232Dual two communication interfaces |
power supply |
220V/110V, 50Hz, 4A |
size |
710mm×210mm×830mm |
Net weight |
90kg |