If an error message appears on the touch screen, search for the following list to identify the error and seek appropriate solutions. Press the 'OK' ke
Product details

Suzhou Lefeng Precision Instrument Co., Ltd(German Japanese Maintenance Department)I specialize inSanfeng from Japan and Mal from GermanyA professional service center for branded measuring tools and instruments. The company has strong technical strength, with multiple engineers from the world's largest measuring tool manufacturing enterprises. The accessories are imported with original packaging, and the measuring laboratory instruments are all imported measuring instruments to solve the problem of poor accuracy after maintenance for customers. The quality service is supported by the metrological quality inspection department, which issues relevant maintenance product inspection reports and cooperates with customersGeneral agent of "Mitutoyo" and "Mahr"Establish long-term stable cooperative relationships. It is one of the most professional measuring instrument maintenance service providers in mainland China.
Professional maintenance: calipers, height gaugesdial indicator, dial gauge, micrometerInternal and External Diameter ScaleThickness gauges, roughness meters, 2D height gauges, tool microscopes, etc.
Service commitment: Provide free on-site pickup, free fault detection, and within 2 days of quick repair service
Quality assurance: The accuracy of the repaired products meets the national precision standards
Warranty period: Non human damage, 3-month warranty for similar faults
Troubleshooting methods for common problems with Japanese Sanfeng roughness tester

If an error message appears on the touch screen, search for the following list to identify the error and seek appropriate solutions. According to the information or close the window, press‘OK“Key or other keys. Then take necessary measures to address the errors.

characteristic/Possible causes and handling methods for error messages

battery cannot be charged●SJ-400I forgot to turn off the current when connecting the AC adapter.

The battery switch is turned off.

The built-in battery is damageddeterioration.

The built-in battery is over discharged.openSJ-400Check the battery icon to ensure that the display icon indicates“Be chargedCharging

Turn on the battery switch

Replace the old battery with a new one.

charge.

Cannot run automatic sleep●RS-232CCommunication is still ongoing.

Perform statistical process.

The communication adapter is still connected.

The automatic sleep function is set toOFFNo problem. existRS-232CAutomatic sleep during communication and statistical processes is ineffective. press“POWER”Just press the key to turn off the power.

The touch screen display is unclear and the contrast adjustment is inappropriate. Open the contrast control button behind the monitor device and adjust the contrast appropriately.

The buttons or touch screen cannot be used●RS-232The mode is still running

The shielding of keys has an effect.turn offRS-232CSwitch.

Turn off the key shield switch.

memory error1Error location: Power on self-test error in internal memory. yourSJ-400Agents or nearest onesMITUTOYOSales office.

memory error2Error location: Internal storage data field error in internal memorySJ-400Agents or nearest onesMITUTOYOSales office.

system error1Error location: Self checking error in input signal line. yourSJ-400Agents or nearest onesMITUTOYOSales office.

SPCerror1Error location: withSPCDevice outputSPCReconnect appropriately due to data exchange errorsSPCConnecting line. Turn on the power and enter the data processor.

print error1Error location: The printer is out of paper and cannot receive printed output. Please add more paper and press againPRINT

print error2Error position: The printer head has an error in adding paper and turning the lever.

print error3Error location: If the battery of the automatic printer is not fully charged40—20%Charging.

Turn off the power and wait for a few seconds. If the remaining battery level exceeds60%--40%Turn on the power. Print again. If the problem persists,

yourSJ-400Agents or nearest onesMITUTOYOSales office.

Storage card error1Error location: When the card is write protected, performing write, formatting, or deletion of the memory card requires removing the memory card from the slot and canceling the write protection,

Storage card error2Error location: The data stored on the storage card exceeds the storage capacity. Insert a new storage card and format it. Implement a new save operation

Storage card error3Error location: A transfer error occurred while reading the storage card. Ensure that the storage card is securely inserted into the slot, and if not, re insert it.

Storage card error4Error location: The storage card has not been formatted. If permission to read and retrieve is required, the storage card may be formatted incorrectly.

Storage card error5Error location: There is no card in the slot during access. Insert a card into the slot and format it if it is new.

LCDThe monitor has nothingThe contrast has not been adjusted properly.

The power remains turned off. When viewed from the front, rotate the contrast adjustment button in the opposite direction to turn on the power. Secondly, adjustLCDContrast ratio.

Driver error1Error location: Driver/Detector device

Within a specific time period, the initial range cannot detect or execute the correct connection driver/Detector device and display device. If the problem persists, yourSJ-400Agents or nearest onesMITUTOYOSales office.

Driver error3Error location: Driver not connected correctly/Correctly connect and drive the detector device/Detector device and display device. If the problem persists, yourSJ-400Agents or nearest onesMITUTOYOSales office.

Limit detection error location: driver/Detector device

Halfway through the measurement within the driving range, the detector device needs to slide and measure the entire length before the end of the detection range has been detected. In the3Page by“Move the measurement start position”Icon or“Return it to the origin”Icon, enter the measurement start positionStart of driving rangeIn order to achieve the measurement length.

Detector error1Error location: detector

Before starting the measurement, the detector is not within the detection range for the workpiece. Correctly set the detector on the workpiece and attach it correctly to the workpiece

Detector device.

Detector error2Error location: detector

When measuring, the detector exceeds the detection range for the workpiece. Correctly set the detector on the workpiece and attach it correctly to the workpiece

Detector device.

Detector error3Error location: detector

The detector is improperly installed on the detector device. The detector is correctly attached to the detector device.

Out of range error display“000”Error location: Measurement range

The detected value exceeds the current measurement range. Use a large measurement range.

Zero position conversion error, incorrect location: probe

The probe position is not within the specified range based on the provided zero position conversion value. Adjust the probe to ensure its position falls within the specified range.

DATerror

Data adjustment platform error location: Tilt adjustment

existDATDuring the measurement period, calculations cannot be performed. The tilt angle between the detector and the workpiece is too large. The tilt of the detector needs to be finely adjusted.

Ra…………umError location: Computing device

Parameter calculation error occurred.

occurRSurface compensation calculation error.Measure again after modifying the measurement conditions.

ifRThe surface compensation has been set toON, during startupRMeasure again after surface compensation.

☆25.65umError location: Computing device

During the measurement period, it was detected that it exceeded the range. Stick a piece in front of the data“☆“Use a large measurement range.

Rz#5.68umError location: Computing device

countRzThe peak valley or peak bottom data is smaller than the specified data.

“#“Display before the calculation result.

Ra****mError location: Computing device

If the calculation result is not obtained.

Rz L-PError location: Computing device

Not enough peak valley or peak bottom data for computationRzperhapsSmBefore the calculation result, it will be displayedL-P.

Calibration Error Location: Calibration

The calibration measurement result exceeds the possible range.

Check the roughness value of the reference standard block for the data displayed on the touch screen. Ensure that the detector is correctly set on the reference standard block.

Condition error1In terms of evaluation criteriaMeasurement conditions1/3and2/3)Evaluation criteria for checking the mismatch between the measured data and the relationship.

Condition error2In terms of evaluation criteriaMeasurement conditions1/3and2/3)Check the measurement conditions for any mismatch between the relationship with the measurement data.

Condition error3Cannot calculateRSurface compensation measurement data cannot be implementedRFace compensation. turn offRFace compensation.

Cannot setRyThe surface roughness standard has been set toJIS01ISO97ANSIperhapsVDA. Standard(JIS2001ISO1997)During the revision, the meaning of parameter symbols was also modified. The old standard versionRychange toRzAvoid usingRyTherefore, according to the standards(JIS2001ISO1997)The instrument cannot be setRyperhapsPyParameters.

If necessaryRySo, change the standard back“JIS‘82”perhaps“JIS‘94”Measurement will only be implemented in any version.