Jenco 6311 Industrial Online pH Meter
characteristic
Standard 1/4 DIN waterproof body design
LCD display, capable of simultaneously displaying pH value, ORP value, and temperature value
PH electrode efficiency display
4-20mA reversible, with isolated current output, user can set output range by themselves
Four sets of pH (ORP) relay control/one set of cleaning relay control, control points/hysteresis width can be set by users themselves
Jenco 6311 Industrial Online pH Meter
Equipped with automatic temperature compensation function, users can choose their own temperature probe (PT1000, TH10K, BALCO3K)
Power off correction value memory function
Using RS-485 communication interface
The product has passed CE certification
?
6311 Industrial Online pH/ORP/Temperature Controller
specifications
pH |
measuring range |
-2.00 ~ 16.00pH |
Resolution/accuracy |
0.01pH/0.1% ± 1 word |
|
Input impedance |
>1013Ω |
|
PH buffer recognition solution |
Users can choose either 4.01, 7.00, 10.01 or 4.00, 6.86, 9.18 |
|
PH electrode zero/slope compensation |
±100mV / ±30 % |
|
ORP |
Measurement range of oxidation-reduction value |
-2500 ~ 2500mV |
Measurement range of relative oxidation-reduction values |
-6499~6499mV |
|
Resolution/accuracy |
1mV/0.1% ± 1 word |
|
temperature |
measuring range |
-10.0 ~ 120.0℃ |
Resolution/accuracy |
0.1℃/±0.3℃ |
|
Temperature sensor |
Thermistor, 10K/25 ℃ or PT1000 or Balco 3K/25 ℃ |
|
4-20mA current output |
current output |
4-20mA reversible, isolated current output |
Edge voltage/load |
500VDC/500Ω |
|
Relay control output |
control mode |
ON/OFF control within the system |
Edge voltage/load |
500VDC/500Ω |
|
other |
key |
All have sound feedback effects |
Protective measures for various settings |
4-digit password protection |
|
External dimensions |
1/4DIN standard body, 96x96x148mm |
|
Dimension |
92x92mm |
|
Waterproof grade |
IP65 |
|
weight |
960g |
|
power supply |
230VAC,50/60Hz |
pH electrode
model |
brief introduction |
ST873 |
High temperature field |
IP-600-1 |
The field of biological fermentation at room temperature and wastewater treatment with high pollution |
IP-600-10 |
General wastewater treatment field |
IP-600-9(TH) |
With temperature compensation, in the field of ordinary wastewater treatment |
ORP | |
IR-500-9N |
General wastewater treatment field |
IR-500-8 |
General wastewater treatment field |
Protective casing
model |
brief introduction |
009A |
1-meter pipe length |
sealed box
model |
brief introduction |
011-200-00 |
Dial type installation |
012-200-00 |
Wall mounted installation |