1、 Product Introduction
JC-310(S)Intelligent dynamic gas distribution instrument (referred to as gas distribution instrument),Gas mixer,The gas proportioner can dilute high concentration standard gas into various low concentration standard gases according to the set dilution ratio, and can calibrate various gas analyzers and their gas sensors. satisfyHJ 57-2017Determination of Sulfur Dioxide in Waste Gas from Fixed Pollution Sources by Potentiometric Electrolysis Method
This gas analyzer is a high-tech product developed to meet the calibration, calibration, and detection needs of gas analysis and testing instruments. This instrument can real-time prepare multi-component mixed gases of different types and concentrations. The target audience includes environmental protection enterprises, biopharmaceuticals, environmental monitoring, coal mining metallurgy, and all other enterprises and units that require mixed gases.
The gas flow control unit of this product adopts a high-precision, corrosion-resistant quality flow controller that meets national standards(MFC,1Stable output gas flow rate, accurate control, and strong anti-interference ability. The internal gas path materials of the system are all made of anti adsorption and anti-corrosion polytetrafluoroethylene pipelines and passivation316LStainless steel joints are suitable for the vast majority of analytical instruments and gas detection applications.
2、 Execution standards
GB/T 5275.7-2014/ISO 6145-7:2009Gas Analysis - Preparation of Calibration Gas Mixtures by Dynamic Volume Method - Part 7: Thermal Mass Flow Controller
HJ 57-2017Determination of Sulfur Dioxide in Waste Gas from Fixed Pollution Sources by Potentiometric Electrolysis Method
3、 Main features
Adopting imported high-precision and corrosion-resistant imported mass flow controllers
All internal pipelines are made of corrosion-resistant and anti adsorption stainless steel or polytetrafluoroethylene materials
Automatically calculate the intake flow rate for each route based on the input standard gas concentration, target gas concentration, and total output flow rate
Equipped with temperature and pressure compensation correction functions, flow control is more accurate
Built in programmable standard gas process input function, can output different concentrations of standard gas according to different time periods according to settings
Embedded high-speed microprocessor control,7Inch high-resolution touch screen display
Built in lithium battery, battery working time is longer than4hour
built-in5Micro fine filtration ensures that the instrument is not contaminated by particulate matter inside
Strong scalability, customizable with different output flow rates and dilution ratios according to user needs, and can increase the number of valve channels
4、 Product parameters
Input channel
Standard configuration3Input channel, MAX flow per channel5L/min(2-6Channel expandable)
Input gas
airN2Or toN2Other standard gases for balancing gases
Input traffic range
10/50/100/200/500/1000/5000ml/min (Customizable)
MAX dilution ratio
1:100
Power Supply
AC220V,50HzorDC24V
Gas path material
passivation316LStainless steel, polytetrafluoroethylene
Output traffic
(0.1-15 )L/min (Customizable)
Accuracy level
1.0
Interface
1/4or1/8(Can be customized according to user needs)
display
7Inch touch screen
weight
about3.5Kg
size
Length x Width x Height:255×340*160