Introduction and Technical Specifications of Ion Chromatography ICS-2100:
ICS-2100 is the first ion chromatography instrument in the history of ion chromatography that integrates both an online eluent generator (RFIC-EG) and a chemical free electrolytic sample pretreatment device (RFIC-ESP), with both isocratic and gradient elution functions. Both the 2mm system chromatography column and the 4mm system chromatography column can work on this instrument. The ICS-2100 ion chromatograph, which uses valve switching and electrolysis water principle, can be used for different pretreatment of various matrix samples without chemical reagents for sample electrolysis. The ICS-2100 system, with its USB connection and mature digital output mode, achieves unparalleled automation. When used in conjunction with the continuous automatic regeneration microfilm suppressor (SRS-300) and the new chameleon software (Chromeleon), it allows users to enjoy convenience while enjoying maximum operational pleasure and convenience.
Technical parameters:
Analyze pumps and conveying devices
Type: Series connected double piston reciprocating pump, microprocessor
Control fixed stroke, variable speed
Structural chemical inertness, non-metallic PEEK pump head and
Flow path, compatible with water with pH ranging from 0-14
Solution and inverse solvent
Compatible with pH 0-14 aqueous solution and reverse phase solvent
Control mode through the front panel or through the chameleon software
Full process control; Can choose TTL or relay
Device switch
Operating pressure 0-35 MPa (0-5000 psi)
Flow rate range 0.00-5.00 mL/min, increasing by 0.01
Quantity, no need to replace pump head
Traffic accuracy<0.1%
Traffic accuracy<0.1%
Pressure fluctuation<1%
Piston seal cleaning dual pump head, continuous cleaning when connected to detergent
Pressure range 0-5000 (PEEK pump head); Can be set
Upper and lower limit alarm function
Rinse solution bottle made of polypropylene, 4L
No need to pressurize the rinsing solution
Injection valve Rheodyne's electrically controlled PEEK valve
Columns 2, 3, 4, and 5 mm i.d., with a maximum length of 250 mm
Analysis column and 50mm protective column
Conductivity detection circuit and flow cell
Type Microprocessing - Digital Signal Control Processor
Linear 1% at 1 mS
Resolution 0.00238 nS
Detection range: Digital format 0-15000 µ S
Filter startup time from 0 to 10 seconds, user selectable
Temperature compensation is fixed at 1.7% for every 1 ℃ pool temperature
Temperature control range of conductivity cell: Environment+7 ℃ to 55 ℃
Passivation of 316 stainless steel electrode in conductivity cell
Conductive cell material, chemically inert polymer material
Conductivity cell volume<1 µ L
Low differential capacity of pool heating exchanger
Maximum pool operating pressure 10 MPa
Suppressor Control Power AES ®, 0-150 mA, incremental 1 mA;
SRS ®, 0-500 mA, incremental 1 mA
The conductivity cell is compatible with methane sulfonic acid
Chameleon software
Compatible with WINDOWS 2000/XP and VISTA systems, the software can
Set automatic power on/off
suppressor
Cation Suppressor Continuous Automatic Regeneration Microplate Suppressor or Column Membrane Mixing
Combined suppressor, with a suppression capacity of 110 micro equivalents
/Minutes
Anion Suppressor Continuous Automatic Regeneration Microplate Suppressor or Column Membrane Mixing
Combined electrolytic suppressor, with a suppression capacity of 200 microns
Equivalent per minute
Suppressor dead volume < 50 µ L
Suppressors do not require the use of peristaltic pumps and have no vulnerable parts
Automatic rinsing solution generator
Concentration range 0.1-100 mM
Flow rate 0.1-3.0 mL/min
Maximum operating pressure 21 MPa (3000 psi)
Maximum solvent concentration Anion: 25% Mathanol
Cation: solvent-free
column oven
Operating temperature range: Environment+5 ℃ to 60 ℃
Temperature stability ± 0.5 ℃
Physical specifications
Power supply 100-240 V AC, 50/60 Hz
(Automatic sensing and adjustment)
Voltage range 90-265 V AC, 47-63 Hz
Instrument operating temperature: 4-40 ℃ (40-104 ° F)
Operating humidity range 5-95% relative humidity, no condensation
The control mode is controlled by chameleon software and can be selected
T T L or delay switch. Two delayed outputs
Out, two TTL outputs, four programmable
Input.
USB communication port with one input and two fixed USB input ports
Leak detection built-in, optical sensor
Dimensions (h × w × d) 56.1 × 22.4 × 53.3 cm;
(22.1×8.8×21 in.)
Weight 24.5 Kg (54 lbs)
AS-DV technical indicators
The sample bottle capacity can hold 50 sample bottles, using 5 mL,
0.5 mL 2 types
Transfer volume 0.1-5.0 mL, with increments of 0.1 mL
Speed control 0.1-5.0 mL/min, with 0.1 mL/min as the
increment
Software control can be achieved through chameleon software
Each bottle can be injected multiple times
Dimensions (h × w × d) 28.7 × 36.3 × 44.5 cm
Weight<8.9 Kg (<195 lbs)
Main features:
Multifunctionality
The ICS-2100 integrated ion chromatography system can be used for the separation of various ions, including anions, cations, organic amines, organic acids, etc.
The RFIC-EG eluent online generation device can convert high-purity deionized water into eluent online.
The RFIC-ESP ion chromatography device for sample pretreatment without chemical reagents can be used to control online water purification devices and other sample pretreatment devices.
The instrument can automatically perform pre-processing on samples with various matrices, including online filtration, matrix elimination, online neutralization, online concentration, etc.
The excellent dual piston reciprocating pump can make the instrument run more smoothly, providing a more ideal operating environment and detection effect.
The flexible flow rate adjustment and wide flow rate range make the instrument compatible with chromatography columns of 2, 3, 4, and 5mm.
The brand new design ensures that the instrument can greatly solve the problem of laboratory area and space.
The operator can also set the main parameters of the instrument through the LCD touch sensitive display panel and monitor the working status of the instrument in real time.
More convenient
● Universal rinse organizer, compatible with 1L, 2L, and 4L glass and plastic rinse bottles
Standard configuration includes the well-known Rheodyne company's electrically controlled PEEK six way valve
The standard configuration includes a manual injection port that conforms to human mechanics
The unique design of the rinsing solution control valve can automatically disconnect the rinsing solution pipeline before the pump, which is more conducive to instrument maintenance
All components are within easy reach, making maintenance and repair of the instrument more convenient
● Equipped with leakage monitoring function, it can detect whether the instrument has any problems in real time
The standard TTL interface allows for easy control of additional pumps and injection valves through the host, providing maximum upgrade space for the instrument
The LCD high brightness color touch LCD display screen is convenient, intuitive, and easy to learn. The ICS2100 panel has an LCD high brightness color touch LCD display screen, and users can directly control and monitor the instrument through the various menus and parameters displayed on the touch screen. High performance pumps are convenient, accurate, and reassuring
The high-tech dual plunger high-pressure pump can provide high-precision, low drift, and pulsation free rinsing solution, ensuring stable baseline and achieving extremely low detection limits; The flow path design of all PEEK materials avoids metal contamination and can withstand high pressure, acid, alkali, and reverse phase organic solvents; The flow rate range of the pump is wide, and it can achieve a flow rate of 0.00-5.00 mL/min without replacing the pump head; Sensitive leakage sensor ensures the accuracy of the analysis process
Relying on sex. The automatic generator for rinsing solution is convenient, worry free, and ensures repeatability. Only one isocratic pump is needed to complete gradient rinsing, greatly saving instrument costs; By providing a parameter, the desired concentration of the rinsing solution can be automatically obtained, which is convenient and fast; The repeatability of the prepared rinsing solution is good, eliminating the errors caused by manual configuration and the impact of errors between different times and operators; Just provide pure water for the instrument, no need for contact testing
Agent, safe and reliable.