Industrial silicate analyzer manufacturerInstrument installation:
1. Installation requirements and precautions:
① The installation position of the analytical instrument should be as close as possible to the sampling point, and the water sample taken should be representative;
② The measured water sample and environmental temperature should be between 5 and 45 ℃, otherwise it will affect the chemical analysis process and thus affect the measurement accuracy;
③ Ensure that the water sample is free of impurities and dirt. If the water quality is not qualified due to maintenance or other reasons, the water sample should be disconnected and the instrument should stop running;
④ There should be no strong electromagnetic fields or vibration sources around the working environment for installing instruments;
⑤ The instrument should be installed in a dry, dust-free, and non corrosive gas environment;
2. Installation and fixing method:
The instrument can be installed on various instrument panels or fixed brackets. When installing on the dial, the hole size is (665 × 405mm) according to the following diagram.
Industrial silicate analyzer manufacturerUsing imported precision measuring solenoid valves instead of traditional pumps, the structure is simple and reliable, with a low failure rate. Adopting imported light sources and receiving components with long lifespan, high stability, and high reliability. Simple, reliable, and intuitive flow path. At the same time, during the color reaction of the instrument, the pathway is switched and thoroughly cleaned, fundamentally solving the problem of cross contamination caused by the alternating measurement of each channel on the measured values. Designed colorimetric cell to eliminate bubble interference.
Technical parameters:
1. Measurement range: 0~20ug/l, 0~2000ug/l, 0-20mg/L optional
2. Accuracy: ± 1% F.S
3. Reproducibility: ± 1% F.S
4. Stability: Drift ≤ ± 1% F.S/24 hours
5. Response time: 12 minutes for initial response
6. Sampling period: about 10 minutes per channel
7. Water sample conditions: flow rate:>100 ml/min
Temperature: 10~45 ℃
Pressure: 10 kPa~100 kPa
8. Environmental conditions: Temperature: 5-45 ℃
Humidity:<85% RH
9. Reagent consumption: Approximately 3 liters per month for each of the three reagents
10. Current output: 0~20mA (can be set arbitrarily within this range, and each channel of the multi-channel instrument can output independently)
11. Alarm output: Relay normally open contact (220V/1A)
12. Power supply: AC220V ± 10% 50HZ
13. Power consumption: ≈ 50VA
14. Dimensions: 720mm (height) x 460mm (width) x 300mm (depth)
15. Hole size: 665mm × 405mm