XRW-300A3-6 Thermal Deformation Vicat Softening Point Temperature Tester
1. XRW-300AThe Type series hot deformation Vicat softening point temperature tester is mainly used for measuring the hot deformation and Vicat softening point temperature of non-metallic materials (such as plastics, rubber, nylon, electrical insulation materials, etc.). It is an ideal testing tool for chemical enterprises, research institutions, colleges and universities, and other departments. Due to the use of high-quality materialsIndustrial control modules, LCD touch screens, and electronic componentsIt constitutes a high-precision data acquisition and control system, ensuring temperature control accuracy and digital sampling accuracy. The instrument operation is simple and convenient, and the instrument runs stably and reliably. This system complies withThe requirements of standards such as ISO75-1993, ISO306-1994, GB1633-2000, GB1634-2000, etc.
2.Instrument characteristics
2.1This machine is controlled by PLC, with a 5-inch high-definition touch screen and AD module for displacement data acquisition. The temperature is collected and processed by a temperature controller, achieving dual control mode of single machine and computer.(Say goodbye to the era of single board computers and enter the era of digital industrial control, which can break away from the dependence on computers).
2.2 Different operating interfaces are used for thermal deformation and Vicat, ensuring accurate, safe, and reliable data of this instrument. It is easy and convenient to operate (both standalone and computer-based).
2.3 According to different standards, the standard deformation of the sample can be set arbitrarily, greatly meeting the requirements of different customers and standards. (Truly achieving timeless models)
2.4 The instrument adopts dual temperature protection functions of software and hardware to ensure the safe use of methyl silicone oil during operation and ensure the safety and reliability of the instrument.
During the single machine test, the operation is convenient and fast, with a high degree of automation. After the test is completed, the results and average values are automatically displayed and heating is stopped. The operation interface and report screen are displayed on one screen during computer control. After the experiment is completed, it automatically ends and stores the report images, which is accurate, reliable, and contains a large amount of information; At the same time, temperature and deformation data at different times can be queried and typed.
2.6 The results of the computer operation test are recorded in the form of images on the hard drive as saved data and can be printed out at any time for future reference and comparison.
3TECHNICAL INDEX
3.1Temperature control range: oil temperature -300 ℃
3.2Temperature resolution: 0.1 ℃
3.3 Temperature control error: ± 1 ℃/60 minutes (120/h)
± 0.5 ℃/60 minutes (50/h)
3.4 Heating rate: 120 ℃/h, 50 ℃/h, others
3.5 Deformation measurement range: 2mm
3.6 Deformation measurement error: ± 0.005mm
3.7 Sample rack quantity: 3
3.8 Test stations: 6
3.9 Temperature sensors: 6
3.10 Heating medium: Methyl silicone oil (below 200 cubic meters, flash point above 300 ℃, customer provided)
3.11 Heating power: 4KW
3.12 Cooling method: Water cooling below 150 ℃, natural cooling above 150 ℃
3.13 Power Supply (Main Power Supply): 220V, 15A, 50Hz
3.14 Weight of Load Rod and Tray: 76 ± 2g