overview
The HY-900A mobile gamma radiation monitoring device is a new type of real-time continuous online measurement and alarm device for gamma radiation. The equipment can be networked at multiple points through wired or wireless means, and can continuously monitor a certain range of areas through multi-point control. The measurement data is transmitted in real time to the control center.
Performance characteristics
- The use of high-speed embedded microprocessors ensures real-time monitoring of data and status.
- The alarm threshold is continuously adjustable (the factory default setting is a high threshold of 25uGy/h and a low threshold of 0.0Gy/h).
- Real time display of the status of each probe: normal/alarm/overload/disconnection.
- Automatically record threshold alarms and store them, without loss after power failure; Threshold alarm records can be manually queried.
Technical Parameter
- Detector: Metal GM tube detector.
- Measurement range: 0.01 μ Sv/h to 3.0mSv/h.
- Sensitivity: 1 μ Sv/h>5CPS.
- Energy response: 48Kev~3Mev ≤± 30%.
Communication method:
Wired: standard RS485 interface, transmission distance up to 800 meters;
Wireless: satellite/public network/radio, etc., optional.
Equipped with a tripod, the deployment height is 0.8-1.5 meters.
HYGP-1123 is a small and high-sensitivity x γ radiation detector, which uses the latest low-power embedded microprocessor as the data processing unit. The detector is composed of a high-sensitivity scintillator and a high range GM counter. The instrument is more sensitive to short-term pulse radiation and can detect short-term (≥ 10ms) X-pulse radiation. It has a wide range of applications, such as environmental monitoring (nuclear safety), radiation health monitoring (disease control, nuclear medicine), homeland security monitoring (entry and exit, customs), public safety monitoring (public safety), nuclear power plants, laboratories, and nuclear technology applications.
Instrument characteristics
▶ Large area backlit color display
▶ The instrument has high sensitivity and can measure environmental background
▶ Automatically save cumulative dose, can be saved for a long time when powered off
Specification
▶ Continuous radiation field measurement range: Dose rate: 0.01 μ Sv/h-200mSv/h
Cumulative dose: 0.00 μ Sv to 9999Sv
▶ Pulse radiation field measurement range: 0.01 μ Sv/h-200mSv/h
▶ Pulse radiation response time: 5ms
▶ Energy range: 20KeV-7MeV
▶ Sensitivity: 1.7cps+1500cps (1 μ Sv/h)