HL-600Portable digital displayleeb hardness tester
■HL-600Portable digital displayThe Leeb hardness tester is the latest product developed by our company, which includes a built-in printerUSBOutput interface.
■ Directly display Richter, Brinell, Roche, Vickers, Shawtensile strengthWaiting for various hardness values,
■ HL-600Type is substitutionHL-80、HL-200、HL-240、HL-300、HL-800、HLN
Product features:
·According to the principle of Leeb hardness measurement, various metal materials can be tested
·One host can be equipped with7Use different types of impact devices and automatically identify the type of impact device,No need to recalibrate during replacement
·Support 'forged steel'(Steel)”Materials, when usedD/DCWhen testing the "forged steel" specimen with a type of impact device, it can be directly readHBValue, no need for manual table lookup. Pre set upper and lower limits for hardness values, automatic alarm for exceeding the range, convenient for users' needs for batch testing ·Adopting a large screen128×64Graphic dot matrix LCD display, rich and intuitive information
·Full Chinese display, menu style operation, simple and convenient operation, withELBacklit display, convenient for use in low light environments ·Maximum storage capacity500Group (number of impacts)1~32)Hardness measurement data, each set of data including single measurement value, average value, measurement date, impact direction, frequency, material, hardness system and other information
·The thermal printer is integrated with the instrument, with fast printing speed and the ability to print inspection reports on site. Equipped with software calibration function
·Built in nickel hydrogen rechargeable battery and charger; Can work continuously for no less than150Hour; Equipped with energy-saving functions such as automatic sleep mode and automatic shutdown ·There is a remaining battery indicator icon on the LCD, which can display the remaining battery level in real time; Equipped with charging process indicators, the charging progress can be monitored at any time haveUSBOutput interface, convenient and fast to connect withPCMachine for data exchange. It can be equipped with microcomputer software, which has rich functions such as transmitting measurement results, storing and managing measurement values, statistical analysis of measurement values, printing measurement reports, and batch setting instrument parameters, meeting higher requirements for quality assurance and management.
·The instrument is compact, portable, highly reliable, suitable for harsh operating environments, and resistant to vibration, impact, and electromagnetic interference.
·External dimensions:
■ Applicable materials:Steel and cast steel, alloy tool steel, stainless steel, gray cast iron, ductile iron, cast aluminum alloy, copper zinc alloy (brass), copper tin alloy (bronze), pure copper forged steel
Main application areas:·Failure analysis of bearings and other components, pressure vessels, steam turbine generators and their equipment
·Heavy workpieces · Installed mechanical or permanently assembled components ·Workpieces with limited experimental space
·Require formal original records of test results and material differentiation in metal material warehouses
·Rapid inspection of multiple measurement points in a large area of large workpieces
■ Technical parameters:
·Measurement range:(170-960)HLD、(17-68.5)HRC、(13-100)HRB、(59-85)HRA、
(19-651)HB、(80-976)HV、(30-100)HS
·Measurement direction: Supports360Omnidirectional and arbitrary direction(Vertical downward, oblique downward, horizontal, oblique upward, vertical upward)
·Hardness standard: LeebHLBu ShiHBLuo'sHRCLuo'sHRBLuo'sHRAVickersHVXiao'sHS
·Indication error: ±4HLD(HLD=760time)
·Show off Display: Dot matrixLCD,128×64Graphic dot matrix LCD
·Thermal printer: capable of printing any number of test results to meet on-site needs
·Printing paper width: (57.5±0.5)mmPrinting paper roll diameter:
·Working voltage:6VNiMH battery pack
·Charging power supply:9V/500mACharging time2.5~3.5hour
·Continuous working time: approximately150Hour (without printing or backlight)
·Communication interface standard:USB1.1
Working conditions:Environmental temperature: Operating temperature -10~+
Relative humidity ≤90%; There is no strong vibration, strong magnetic field, corrosive medium, or severe dust in the surrounding environment.
■ Operating instructions:After booting up, it will automatically enter the main display interface:
·Battery level: Display remaining battery level ·Impact direction: Current impact direction
·Average value: appears when the average value is displayed after reaching the set number of impacts
·Hardness system: The hardness system of the current measured value
·Measurement value: current single measurement value (no average prompt), current average value (with average prompt)
·Show off Display: indicates exceeding the conversion or measurement range, indicates falling below the conversion or measurement range
·Materials Material: The currently set material
·Impact frequency: Display the completed impact frequency during measurement, display the set impact frequency when setting the impact frequency with the frequency shortcut key, and display the corresponding number of times for a single measurement value when browsing.
■USBData processing software:
·Storage: Store the measured data ·Call out: call out data from the host
·Printing: Print and output data,A4interface ·Over limit: Display of over limit after setting tolerance limit
·Clear: Clear measurement data ·Connection: Set up connection with the host ·Disconnect: Set to disconnect from the host
·Download: Download data locally to achieve massive storage · Settings: Parameter settings · Help: Instant help function
Selection of impact device:
HL-600The standard sensor isDType, suitable for most occasions.
HL-600DCThe type of sensor isDCType, suitable for situations with limited space, such as bucket shaped inner walls or inside assembled machinery.
HL-600D+15The type of sensor isD+15Type, suitable for use in grooves or concave surfaces.
HL
HL-600EThe type of sensor isEType, made of diamond material, suitable for work with extremely high hardness.
HL
HL
Various material measurement tables:
material |
Hardness system |
Impact device |
|||||
D/DC |
D+15 |
C |
G |
E |
DL |
||
Steel and cast steel Steel and cast steel |
HRC |
17.9~68.5 |
19.3~67.9 |
20.0~69.5 |
|
22.4~70.7 |
20.6~68.2 |
HRB |
59.6~99.6 |
|
|
47.7~99.9 |
|
37.0~99.9 |
|
HRA |
59.1~85.8 |
|
|
|
61.7~88.0 |
|
|
HB |
127~651 |
80~638 |
80~683 |
90~646 |
83~663 |
81~646 |
|
HV |
83~976 |
80~937 |
80~996 |
|
84~1042 |
80~950 |
|
HS |
32.2~99.5 |
33.3~99.3 |
31.8~102.1 |
|
35.8~102.6 |
30.6~96.8 |
|
Steelforged steel |
HB |
143~650 |
|
|
|
|
|
CWT、ST alloy tool steel |
HRC |
20.4~67.1 |
19.8~68.2 |
20.7~68.2 |
|
22.6~70.2 |
|
HV |
80~898 |
80~935 |
100~941 |
|
82~1009 |
|
|
Stainless steel stainless steel |
HRB |
46.5~101.7 |
|
|
|
|
|
HB |
85~655 |
|
|
|
|
|
|
HV |
85~802 |
|
|
|
|
|
|
GC. IRON Grey cast iron |
HRC |
|
|
|
|
|
|
HB |
93~334 |
|
|
92~326 |
|
|
|
HV |
|
|
|
|
|
|
|
NC、IRON nodular cast iron |
HRC |
|
|
|
|
|
|
HB |
131~387 |
|
|
127~364 |
|
|
|
HV |
|
|
|
|
|
|
|
C.ALUM Cast aluminum alloy |
HB |
19~164 |
|
23~210 |
32~168 |
|
|
HRB |
23.8~84.6 |
|
22.7~85.0 |
23.8~85.5 |
|
|
|
BRASSCopper-Zinc alloys (Brass) |
HB |
40~173 |
|
|
|
|
|
HRB |
13.5~95.3 |
|
|
|
|
|
|
BRONZECopper tin alloy (bronze) |
HB |
60~290 |
|
|
|
|
|
COPPERPure copper |
HB |
45~315 |
|
|
|
|
|
Introduction Table of Technical Parameters and Applicability of Irregular Impact Device:
Irregular impact device |
DC(D)/DL |
D+15 |
C |
G |
E |
|
Impact energy Impact mass |
11mJ |
11mJ |
2.7mJ |
90mJ |
11mJ |
|
Ball head hardness Ball head diameter Ball head material |
1600HV tungsten carbide |
1600HV tungsten carbide |
1600HV tungsten carbide |
1600HV tungsten carbide |
5000HV diamond |
|
Diameter of impact device Length of impact device Weight of impact device |
86(147)/ |
|
|
|
|
|
Maximum hardness of the specimen |
940HV |
940HV |
1000HV |
650HB |
1200HV |
|
Average surface roughness of the specimenRa: |
1.6μm |
1.6μm |
0.4μm |
6.3μm |
1.6μm |
|
Minimum weight of test piece Can be directly measured and requires stable support Need for dense coupling |
> 2~ 0.05~ |
> 2~ 0.05~ |
> 0.5~ 0.02~ |
> 5~ 0.5~ |
> 2~ 0.05~ |
|
Minimum thickness of test piece Minimum depth of dense coupled hardening layer |
≥ |
≥ |
≥ |
≥ |
≥ |
|
Ball indentation size |
||||||
hardness300HVtime |
Indentation diameter Indentation depth |
24μm |
24μm |
12μm |
53μm |
24μm |
hardness600HVtime |
Indentation diameter Indentation depth |
17μm |
17μm |
8μm |
41μm |
17μm |
hardness800HVtime |
Indentation diameter Indentation depth |
10μm |
10μm |
7μm |
-- -- |
10μm |
Scope of application of impact device |
DType used for routine measurementDCType measuring hole or cylindrical tubeDLMeasuring slender narrow slots or holes; |
■ Standard configuration:
Instrument host |
1tower |
printing paper |
1roll up |
DType impact device |
1only |
Delivery Attached Documentation |
1share |
Standard Leeb hardness block |
1piece |
ABSInstrument box |
1only |
nylon brushA |
1only |
Software CD |
1cover |
Small support ring |
1only |
Communication cables |
1strip |
Power adapter (charger) |
1only |
Accompanying documents |
1cover |
■ User selection:
Irregular impact device |
7species |
Impact ball head |
2species |
Irregular support ring |
12individual |
Replace the impact ball head tool |
1pay |
■ After sales service commitment:
1. 3Free replacement for quality issues within one month.
2. Free two-year warranty (excluding impact device), lifetime maintenance