The HK-1351 dissolved oxygen analyzer is a high-performance online measuring instrument newly launched by the company. The instrument adopts a dissolved oxygen sensor with a measurement accuracy of up to micrograms, which has high sensitivity, fast reaction speed, and can continuously measure online. Instruments can be applied in industries such as power, chemical, metallurgical, food, and pharmaceutical.
◆Independently developed electrode with small zero drift, low residual current, fast response speed, long service life, and simple maintenance
◆Temperature measurement can be configured as platinum resistance, thermistor, or manually set
◆Equipped with an embedded operating systemARM9Processor, secure and stable
◆Internal two independent power modules improve reliability and reduce on-site interference
◆Durable cast aluminum casing, suitable for various installation methods
◆Two channels can be configured with multiple modes of isolation current output function, supporting maximum800Ωload
◆Adjustable alarm delay function to reduce misoperation caused by on-site interference
◆Can be usedSDCard implementation for on-site upgrade or data retrieval
◆Large capacity file system for storing large capacity historical data
◆Four relay outputs, configurable with high and low alarm outputs, and electrode cleaning function
◆quarantineRS485Interface standardsModbusThe data bus of the protocol realizes digital fieldbus technology
Measurement indicators |
dissolved oxygen |
|
display |
480 x 272Dot matrix color screen LCD, can choose to display Chinese or English |
|
Measuring electrode |
dissolved oxygen/Temperature composite electrode |
|
measuring range |
0~20)mg/L |
|
temperature compensation |
0~60)℃ |
|
resolving power |
0.1μg/L;0.1℃ |
|
response time |
≤30s |
|
water kind want seek |
Water sample temperature |
(5~50)℃ |
Water sample pressure |
≤0.3MPa |
|
Water sample flow rate |
(50~300)mL/min |
|
Solid particles |
Diameter less than5μmNo gelatinous substance |
|
Relay output |
Four relay outputs,Configurable alarm outputs for high and low alarms, as well as electrode cleaning function(The cleaning method and cleaning time can be programmed |
|
Current output |
(0.00~20.00) mA, dual channel800ΩLoad, error0.1%FS |
|
Communication method |
RS485、ProfibusorHartCommunication (user selectable) |
|
data storage |
≥10Ten thousand points |
|
environment condition |
Temperature:(5~40)℃Relative humidity:≤95% |
|
Power requirements |
AC(100~240)V,(50~60)Hz |
|
Appearance size |
(146×146)mm |
|
Installation method |
Embedded, wall mounted, and pipe mounted |
|
weight |
≤3kg(Secondary instrument) |
● host 1platform
● Dissolved oxygen electrode 1branch
● Online dissolved oxygen electrode flow pool 1set
● Dissolved oxygen electrode membrane 2slice
● Dissolved oxygen electrode electrolyte 1bottle