Professional Instrument Equipment and Testing Solution Supplier - JETYOO Industrial&Jianyou (Shanghai) Measuring Instrument Co., Ltd, a former Agilent product technical support engineer-Jian JETKEITHLEY Product Technology Application Engineer-RongyooA trade agency co founded in 2011 with a focus on technical support, aiming to break through the old and establish a new one! Fill the gap of most imported instrument and equipment manufacturers only setting up sales points in China, with weak or no technical support, and agent distributors only specializing in sales, not professional in pre-sales technical Q&A/testing plans, and not professional in after-sales training/maintenance calibration. Our technical sales engineers all have a bachelor's degree or above and over 10 years of experience in the testing industry. With our advanced and unique business philosophy, we specialize in providing instruments and equipment, testing solutions, technical training, and maintenance and measurement services to Chinese users. We are a leading company in the East China region of ShanghaiTechnology orientedA comprehensive service provider for instruments and equipment.
HIOKI BT3564 Japanese Nippon Battery Tester
Can correspond to direct measurement of 1000V
Suitable for production lines of high-voltage batteries for EV and PHEV
Internal resistance measurement 0.1m Ω~3k Ω (total package resistance/bus resistance)
Equipped with high-voltage test fireproof flower function
Analog output function
Measurement probe capable of handling 1000V and high-voltage battery packs (optional)
Test lines are not standard. Please purchase a testing line according to the measurement purpose. As a maintenance component, the male connector (system end) for EXT I/O can be used.
Maximum input voltage 1000V. High voltage resistant battery tester for EV and PHEV
BT3564 is a battery tester suitable for inspecting the entry and exit of various battery packs, from high-voltage EVs and PHEVs to residential batteries.
The maximum input voltage is 1000V, which can measure both internal resistance and battery voltage simultaneously.
It is safe and easy to measure high-voltage battery packs using probes corresponding to 1000V
If the L2110 probe corresponding to 1000V in the options is used, it can safely and easily measure between the wide terminals, recessed electrodes, exposed busbars, etc. of the high-voltage battery pack.
High voltage testing fireproof flower function, making measurement safer
Spark discharge is a common phenomenon that occurs during high-voltage measurements.
BT3564 controls the current flowing through the electrodes of the battery pack and reduces the occurrence of spark discharge.
Using analog output function to achieve long-term monitoring
Equipped with analog output function for measuring resistance values. By combining a recorder and a data acquisition device, it is possible to monitor the total resistance value for a long time.
HIOKI BT3564 Japanese Nippon Battery TesterBasic parameters (accuracy guarantee time of 1 year, adjusted accuracy guarantee time of 1 year)
Maximum input voltage rated input voltage: DC ± 1000 V, maximum rated voltage to ground: DC 1000 V
Resistance measurement range 3 m Ω (maximum display 3.1000 m Ω, resolution 0.1 µ Ω)~3 k Ω range (maximum display 3.1000 k Ω, resolution 0.1 Ω), 7-speed switching
Measurement accuracy: ± 0.5% rdg. ± 5 dgt. (30 m Ω~3 k Ω range), ± 0.5% rdg. ± 10 dgt. (3 m Ω range)
Measurement source frequency: 1 kHz ± 0.2 Hz. Test current: 100 mA (3 m Ω range),~10 µ A (3.000 k Ω range), open circuit terminal voltage: 25 V peak (3/30 m Ω range), 7 V peak (300 m Ω range), 4 V peak (3 Ω~3.000 k Ω range)
Voltage measurement range: DC 10 V (resolution 10 µ V) to DC 1000 V (resolution 1mV), with 3-speed switching
Measurement accuracy: ± 0.01% rdg. ± 3 dgt
Display resistance ± 31000, voltage ± 999999 (± 999999 or ± 110000 at only 1000V) LED
Sampling time FAST, MEDIUM, SLOW
Comparison judgment: Hi/IN/Lo (independent judgment of resistance and voltage), comprehensive judgment (AND operation of PASS/FAIL, resistance judgment result, and voltage judgment result), screen display, buzzing sound, external I/O output (open collector, 35 V, DC50 mA max)
Simulated output resistance measurement value (displayed value DC 0 V~3.1 V)
Interface EXT I/O, RS-232C, printer (via RS-232C), GP-IB
Power supply AC100~240 V, 50/60 Hz, 30 VA Max
Volume and weight 215W × 80H × 329D mm, 2.6 kg
Attachment user manual x 1, power cord x 1, precautions for use x 1
Regarding probe length
A: 2 Forks~Probe to Probe B: Probe Length L: Total Length Test Line A (Corresponding to High Voltage Battery 1000V)
Needle shaped test line L2110, A: 750mm, B: 215mm, L: 1880mm, high-voltage battery measurement, DC 1000V
Needle 9772-909772 for tip exchange (1 piece)
Needle type test line L2100, A: 300mm, B: 172mm, L: 1400mm
Used for measuring high voltage batteries, maximum DC 600V, only applicable to BT3563 and BT3562
option
Zero adjustment board Z5038, used for L2110, L2020, 9465-10, 9772, cannot be used for zero adjustment of needle test lines 9770, 9771.
Test line B (for measuring batteries below 60V)
Needle type testing line 9770, A:260mm B:140mm L:850mm DC 70V, Single axis telescopic measuring with a diameter of 1.8mm for small electrodes
Needle type testing line 9771, A:260mm B:138mm L:850mm DC 70V, Needle type for measuring small holes and small objects on the substrate, with a parallel measurement of 0.2mm
Test line C (for measuring batteries below 60V)
Clamp type test line L2107, A: 130 mm, B: 83 mm, L: 1.1 m
4-terminal test wire 9453, A:280 mm, B:118 mm, L:1.36m
Large clamp type test line 9467, with a distance of 30cm between the two forks and the probe, and a length of 1.1m (for non CE marked products)
PC communication
RS-232C connection cable 9637, 9pin-9pin, 1.8m long
GP-IB connection wire 9151-02, 2m long