The HGBS221 closed flash point tester is used to determine the closed flash point of fuel oil, highly volatile light oil products, and other highly volatile flammable liquids. The testing process is fully automated, with automatic correction and calculation of atmospheric pressure, large screen LCD display, Chinese graphical operation interface, and automatic saving and processing of test data, making the operation more convenient, clear, and intuitive. It is an ideal solution for automatic high-precision testing of flash points in the power, petroleum, chemical, and fragrance industries, as well as in testing laboratories.
Main technical parameters
Applicable standards:GB/T 261-2008 ISO 2719-2002
Temperature measurement range: room temperature~300℃
resolving power:0.1℃
Temperature measurement accuracy: room temperature~250℃,≤±0.5℃;250℃~300℃,≤±1℃
Repeatability: ≤4.5℃
Reproducibility: Flash point at110℃ or below110At ℃, ≤6℃; Flash point higher than110At ℃, ≤10℃
Environmental temperature:5℃~40℃
ambient humidity:≤85%(No condensation)
Power Supply:AC220V±10% 50Hz±5%
Power: ≤300W
External dimensions:400×270×450mm
main features
The instrument adopts a modular design, with one host capable of carrying one slave, and the testing module can be flexibly configured
Adopting a graphical user interface with data storage and query functions
Flash point detection method, using differential detection method
■ The ignition source can be replaced by gas or electric sources
Automatic brightness control for electronic ignition devices, ensuring stable brightness throughout the lifespan of the ignition device