characteristic:
-The sterilization temperature can be set to a maximum of 138 degrees Celsius
-Using PT1000 sensor, temperature control is more precise, with a display accuracy of 0.1 degrees
-Two water level monitoring functions: capacitive water level sensor and temperature based water level sensor
-Equipped with a 5-level password setting function, it can be used for data tracing in pharmaceutical factories, meets GMP and FDA certification requirements, and can print FO values
-A pressure gauge that can be configured to display negative pressure up to a maximum of 0.5. Overcurrent and short circuit protection system
-Standard optimization to prevent dry burning function, over temperature protection system, over pressure protection, safety valve
-Open the lid to stop heating function
-Real time data saving function, memory storage of 60 programs, programmable sterilization programs in various modes
-Voice prompt function
-After sterilization is completed, the exhaust valve can exhaust steam at the set six level exhaust speed, and the exhaust speed can be manually adjusted at any time during the exhaust process
-Freertos Embedded Operating System
-Standard USB flash drive for real-time data recording
-Equipped with automatic detection to verify the functional relationship between temperature and pressure, ensuring the achievement of true saturated steam state
-Automatically detect cold air emissions, ensure steam internal discharge function, and guarantee pollution-free environment
-Boiling point temperature can be adjusted according to altitude
-The exhaust temperature can be adjusted within the boiling point temperature range of 0-25 degrees Celsius
-Dual temperature sensors that can detect air and water temperatures
-One item sensor hole on the pot lid and two verification holes reserved on the tank body for easy verification and connection
-Self sensing pressure interlock device, closed lid inspection system, leakage protection device.
|
![]() |
Product advantages |
|
Inner row After cooling, steam is discharged through a professional condenser to achieve true internal discharge and eliminate environmental pollution
cooling fan After sterilization, it automatically starts, and the optimized three-dimensional condensing fan blows air to the entire inner container, shortening the cooling time
Automatic water replenishment
Drying function |
Easy to operate:
|
Display accuracy:
0.1 Sterilization Stage Graphic Display The opening temperature can be adjusted according to customer requirements below 95 ℃ |
|
Internal electrochemical polishing:
After sandblasting, the inner liner undergoes electrochemical polishing to reduce the area of bacterial adhesion, prevent dirt residue, and facilitate user cleaning and maintenance |
|
Dry burning protection:
Standard low water level sensor and heating rate software program, effectively preventing dry burning |
|
Efficient filter:
Equipped with a 0.2um HEPA university filter, it has the function of filtering bacteria and truly achieves zero pollution |
Numerous advantages: Sterilization is safer and more thorough
STIK provides you with high-quality sterilizers, always adhering to the principle of "safety first"
secure
Pressure Chain Institution -Power on, sterilization start, real-time micro position switch during sterilization process Once there is an abnormal alarm at the location of the monitoring chain organization, the machine will be shut down to ensure safety. -As long as there is pressure inside the chamber (with a pressure sensor), the lid cannot be opened. -The lid of the pot cannot be opened when there is a sudden power outage during the sterilization process. -After sterilization, the opening temperature can be set arbitrarily within the range where the temperature inside the chamber is below 90 ℃. |
-Double overpressure protection -Double over temperature protection -Dual exhaust solenoid valve -Prevent dry burning -Pressure sensor -Leakage protector -Fully automatic intelligent control system |
|
![]() |
Capacity (L)/Model | Diameter 370 standard type | 450 diameter standard type | Automatic water replenishment type | drying type | Deodorizing type |
54 liters | IMJ-54A | - | - | - | - |
78 liters | IMJ-78A | - | - | - | - |
65 liters | - | IMJ-65A | IMJ-65W | IMJ-65D | IMJ-65S |
85 liters | - | IMJ-85A | IMJ-85W | IMJ-85D | IMJ-85S |
100 liters | - | IMJ-100A | IMJ-100W | IMJ-100D | IMJ-100S |
120 liters | - | IMJ-120A | IMJ-120W | IMJ-120D | IMJ-120S |
Inner liner electrolytic polishing | have | have | have | have | have |
Inner liner material thickness | 2.5 | 3 | 3 | 3 | 3 |
Display Accuracy | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
Maximum sterilization setting temperature | 138 | 138 | 138 | 138 | 138 |
gas cylinder | side place | side place | built-in | built-in | built-in |
Sterilization range (degree) | 105-138 | ||||
Dissolving temperature (degrees) | 50-100 | ||||
Insulation temperature (degrees) | 40-60 | ||||
set time | Sterilization and melting: 0-9999 minutes; Insulation time: 1-72 hours; Appointment sterilization time: 0-99 hours | ||||
Pressure display range | -0.1-0.5mpa | ||||
Exhaust method | Full exhaust, micro exhaust, no exhaust, intelligent multi-stage exhaust | ||||
Safety device | Pressure safety valve, preventing dry burning, door lock safety interlock device, chamber cover closure detection system, overvoltage detection, overtemperature detection, leakage and short circuit detection | ||||
Automatic water replenishment | not have | not have | have | have | have |
dry | not have | not have | not have | have | not have |
Internal exhaust function | not have | not have | not have | not have | have |
Deodorization system | not have | not have | not have | not have | have |
cooling fan | Horizontal arrangement | Vertical layout | Vertical layout | Vertical layout | Vertical layout |
Water level sensing | have | have | have | have | have |
Pressure sensor | have | have | have | have | have |
USB | have | have | have | have | have |
Opening method of pot lid | One click lifting lid | ||||
operation temperature | 5-45℃ |
Rongsheng (liters) | 54 | 78 | 65 | 85 | 100 | 120 |
rated power | 2800W | 3500W | 3500W | 4800W | 4800W | 4800W |
Sterilized gallbladder size (diameter * height) | φ370*508 | φ370*730 | φ450*465 | φ450*605 | φ450*710 | φ450*845 |
Basket size (W * D * H) mm | φ340*190*2 |
φ340*230 φ340*350 |
φ430*190*2 | φ430*230*2 |
φ430*230 φ430*350 |
φ430*350*2 |
Product weight | 79kg | 120kg | 110kg | 145kg | 180kg | 216kg |
Dimensions (W * D * H) mm | 612*520*840 | 612*520*1070 | 815*660*985 | 815*660*1125 | 815*660*1230 | 815*660*1365 |
Working power supply | 220V±10%,16A,50HZ/60HZ | 220V±10%,32A,50HZ/60HZ |
-The testing technical data is only applicable to the test results under no-load conditions of the equipment and at an ambient temperature of 25 degrees Celsius!
-Product appearance and parameter changes will not be notified separately. The product appearance may deviate due to photography, printing, and other reasons. We apologize for any inconvenience caused!
-Printer (printing content: pot pressure during sterilization, sterilization temperature and time)
-Item sensor
-3Q certification
-Full internal row