FMI51-A1AGDJB2A1A/E+HSanitary process connection Tri Clamp ISO02852 clamp, seamless milk joint installed flush
Sensor length Total length: 6m (20ft) Shielding section length: Maximum 2m (7ft)
Maximum measurement distance 0.1... 4.0 m (0.3... 13 ft)
Communication 4... 20mA HART
PFM certificate and certification ATEX, FM, CSA, IEC Ex, TIIS, INMETRO, NE
No calibration required (factory pre-set). When measuring media with conductivity of 100 μ S/cm or higher, calibration is not required
The process liquid receiving components are made of FDA certified materials that are corrosion-resistant
Menu guided on-site settings through plain text display (optional)
Two level overvoltage protection
application area
The instrument is not affected by the dielectric constant (dc), the conductivity of the liquid is 100 μ S/cm, and it allows measurement of multiple liquids without the need for recalibration.
Process connection: flange and thread
Process pressure: -1...+100 bar (+1.450 psi)
Measurement range: 0.1... 4.0 m (0.3... 13 ft)
Temperature: -80...+200 ° C (-112...+392 ° F)
FMI51-A1AGDJB2A1A/E+HInternational explosion-proof protection certification, overflow protection WHG, SIL, sanitary certification, classification certification
Features and specifications
Continuous liquid level measurement
measuring principle
Capacitive level gauge
Features/Applications
Fully insulated rod probe, suitable for standard and harsh working conditions (temperature, process, adhesion)
Interface measurement
Liquid/liquid interface with lotion layer
characteristic
Shield segment length
Factory calibration
short response time
Power/Communication
12...36 DC HART
PFM
measurement accuracy
Repeatability is 0.1%
Linear error in measurement of conductive liquids
<0.25%
ambient temperature
-50...+70°C