The instrument has a fluoride ion concentration value[F-]In addition to display function, it also haspFValue display.
Main technical indicators:
1Measurement range:a)pFValue:(1.00~6.00)pF;
b) mVValue:(-1800~1800)mV.
2Resolution:0.01 pF.
3Basic error of electronic unit: ±0.02pF±1One word.
4Basic instrument error: ±0.05pF.
5Instrument repeatability error: not greater than ±0.03pF±1One word.
6Temperature compensation range of the tested solution:(5.0~ 45.0)℃
7 Stability of electronic unit: ±0.02pF /3h.
8 Power supply:AC 220V±22V; Frequency50±1HZ
9 External dimensions:1×b×h,mm:200×210×70