Measurement range: 0.04 to 4.0 mg/L (ppm) CI2
Adopting unique SMD microcomputer control technology, high-quality narrow band filter, temperature compensated LED optical system, built-in reaction timer, ensuring the results between sample measurement and user *, compact overall design, waterproof sample pool, ensuring fast and accurate measurement, good reproducibility, with calibration function and built-in software adjustment function, applied in scientific research, metrology for sample detection and identification.
md100_powderpacks.jpg
Chlorine is a commonly used water disinfectant with a wide range of applications, from the treatment of drinking water and wastewater to the sanitation treatment of swimming pools and mineral resorts, as well as disinfection and sterilization in food processing.
ET7000 waterproof residual chlorine total chlorine (Cl2) concentration analyzer, using unique SMD microcomputer control technology, high-performance narrowband filter and temperature compensated LED light source system; The overall design is compact, waterproof sample pool, and the measurement is fast, accurate, and reliable;
Applied for measuring water treatment, wastewater/drinking water, cooling water/boiler water, swimming pool water, etc.
Performance Features Highlights
New waterproof design, exquisite and portable, background light display screen
▷ Good GLP function, automatic shutdown setting
16 sets of data storage, which can be accessed and read at any time, making it convenient for on-site data organization and recording
md100_drops.jpg md100_irim.jpg
Factory and user-defined dual calibration function, automatically remembers the last zero calibration, no need for frequent zero calibration
Unique infrared transmission design, IRiM infrared data transmitter and computer connection facilitate data transmission
275650_reference standard kit_chlor_0_2_und_1_0_mg_l_case_mb2.jpg
Standard verification kit, solving your worries
The identification standard verification kit facilitates users to regularly test instrument performance
Waterproof sample pool design, prefabricated matching water agent,
Stability and validity up to 1-2 years, accurate and reliable measurement results
Technical Specifications
ET7000 residual chlorine [free chlorine] - total chlorine
measurement range
0.02 to 4.0 mg/L(ppm)CI2
resolution
0.01 mg/L
Accuracy @ 25 ° C (77 ° F)
±0.05 mg/L(0 to 1 mg/l)、±0.10 mg/L(1 to 2 mg/l)
±0.20 mg/L(2 to 3 mg/l)、±0.30 mg/L(3 to 4 mg/l)
optical system
Waterproof sample pool, temperature compensated LED optical system;
Wavelength accuracy: ± 1 nm; Resolution: 0.01A, Accuracy: 3% F.S. @ 20 to 25 ° C
Method Standard
Refer to USEPA method 330.5 and standard method 4500-Cl G
calibration mode
Automatic zero point calibration, factory calibration, and user-defined dual calibration modes
data management
16 sets of data storage, which can be accessed at any time; Cooperate with ET214050 IRiM infrared data transmitter for convenient data transmission with PC
Power mode
4 x 1.5AAA battery, automatically shuts down after 10 minutes without operation, LCD backlight
Applicable Environment
5 to 40 ° C, RH: 30 to 90% non condensing
Dimensions & Weight
Host size: 155 × 75 × 35mm, host weight: 260g
certification standard
CE certification, IP68 (under 0.1m conditions, 1 hour)
Recommend purchasing accessories
ET197620 custom glass colorimetric dish, specification: H48mm × ∅ 24mm, 12 sets/group
ET471056 Customized Residual Chlorine (Free Chlorine) - Total Chlorine DPD3 Reagent Type: Water Agent, Specification: 15mL × 3
ET214050 customized dedicated IRiM infrared data transmitter, USB interface, suitable for data transmission to a computer for data, management, and analysis
ET197635 customized glass colorimetric dish cleaning cloth, packaging: 4 pieces/group
ET275650 customized chlorine standard group, standard values: 0.2 and 1.00 mg/L (ppm) CI2