VIP member
ET6800 microcomputer chloride (Cl) concentration analyzer
ET6800 microcomputer chloride (CI) concentration analyzer
Product details

ET6800 microcomputer chloride (CI) concentration analyzer
Range: 0.5 to 25 mg/L, 5 to 250 mg/L CI determination, infrared data transmission function, 16 sets of data storage, can be accessed at any time, GLP management function, factory calibration and user-defined calibration dual-mode, waterproof design, suitable for rapid sample analysis and measurement in laboratories and on-site.
The ET6800 chloride concentration analyzer adopts SMD microcomputer control technology, high-quality narrow band filter, temperature compensated LED optical system, compact overall design, waterproof sample pool, ensuring fast and accurate measurement with good reproducibility. It has calibration function and built-in software adjustment function, and is used for scientific research, measurement, sample testing, and identification.
Standard configurationThe ET6800 chloride concentration analyzer adopts SMD microcomputer control technology, high-quality narrow band filter, temperature compensated LED optical system, compact overall design, waterproof sample pool, ensuring fast and accurate measurement with good reproducibility. It has calibration function and built-in software adjustment function, and is used for scientific research, measurement, sample testing, and identification.
● Exquisite and portable, easy to operate, precise and fast measurement
● Background light display screen, GLP function, automatic shutdown setting
● Memorize usage methods, quickly search for all methods, and ensure standard consistency
16 sets of data storage, which can be accessed and read at any time, making it convenient for on-site data organization and recording
Infrared transmission design, IRiM infrared transmitter (ET214050) and computer connection
● Factory and user-defined dual calibration function, automatically remembers the last zero calibration when the instrument is turned off, eliminating the need to calibrate the instrument zero every time it is measured, ensuring accurate and reliable measurement
Waterproof sample pool design, prefabricated supporting reagents, more reasonable operation and maintenance
● Applied to water treatment, wastewater/drinking water, cooling water/boiler water, swimming pool water, etc
Technical specifications for ET6800 chloride concentration analyzer | ||
Host model | MD100 | |
range | 0.5-25mg/L Cl | 5-250mg/L Cl |
accuracy | ±2.5mg/L | ±25mg/L |
resolution | 0.1mg/L | 1mg/L |
Reagent type | ET517741 (ET515910+ET515920) Chloride Reagent (Tablets 100) | |
measuring method | Silver nitrate/turbidity colorimetric method | |
optical system | 530nm, Waterproof sample pool with temperature compensated LED optical system, wavelength accuracy ± 1nm | |
Photometric index | Resolution: 0.01A, Accuracy: 3% F.S | |
Calibration Mode | Automatic zero point calibration, using SMD microprocessing technology, with factory calibration and user-defined calibration modes | |
data management | 16 sets of measurement data are automatically stored using the IRiM infrared data transmission module | |
Power supply mode | 4X1.5AAA battery with automatic shutdown and power-saving function (10 minutes) | |
Usage environment | 5-40 ° C; RH: 30-90% non condensing | |
Dimension | 155×75×35mm/260g |
MD100 host, standard ET517741 (ET515910 reagent+ET515920 reagent) chloride reagent (tablet, 100), ET197620 (Ø 24mm) dedicated glass colorimetric dish (2 pieces), ET197635 colorimetric dish cleaning cloth (2 pieces), Chinese and English user manual, instrument specific carrying case
Online inquiry