The ECA510A absolute pressure transmitter and ECA530A pressure transmitter are used to measure the pressure of liquids, gases, and vapors, and convert them into 4-20mADC current signals for output. ECA510A and ECA530 can also communicate with each other through the HART275 handheld controller for setting, monitoring, etc.
●Functional specifications
Output 2-wire system, 4-20mA DC output, digital communication, programmable linear or square root output mode, HART protocol loaded on 4-20mA DC signal.
supply voltage
10.5~42V DC (working state)
16.4~42V DC (digital communication)
16.4~30V DC (intrinsic safety type)
Load (when output signal codes are D and E)
0-1335 Ω working state
250-600 Ω digital communication HART
Communication Distance
When using multi-core twisted pair cables, the maximum communication distance can reach 2km. The communication distance varies depending on the cable type
Damping time constant
The sum of damping time constants for amplifier components and membrane enclosures. The damping time constant of the amplifier components can be adjusted within the range of 0.2 to 64 seconds. ambient temperature
-40~85℃(-40~248°F)
-30-80 ℃ (-22-176 ° F) [with LCD meter head]
Liquid contact temperature
-40~120℃(-40~248°F)
Reference accuracy for range adjustment
(including linearity, hysteresis, and repeatability starting from zero)
±0.075%
stability
± 0.1% range upper limit/12 months