Project Content
DL-8001 Insulation Oil Gas Chromatography Analyzer
Temperature control technical indicators
±0.1 Technical specifications of detector
Sensitivity (TCD)
Z
Small detection concentration (ppm)
≤5.0×10-12g/s (n-C16) ≤0.1mv /30min(1.0×10-12A/30min) CO≤0.5 CO2≤2 O2≤10 Other indicators and usage conditions
project
Temperature control range (℃)
Temperature control accuracy (℃)
Temperature increment (℃)
Temperature gradient (%)
Converter temperature rise time (room temperature~380 ℃)
column oven
40~399
1
±1
≤10
thermal conductivity cell
50~399
±0.03
other regions
50~399
±0.2
name
Sensitivity (FID)
noise
drift
linear range
Hydrogen flame ionization detector (FID)
≤0.02mV(2.0×10-13A)
≥106
C2H2≤0.5
Thermal conductivity detector (TCD)
≥5000mVmL/mg(n-C16)
≤0.1
≤0.5
≥104
H2≤10
parameter
index
supply voltage
AC220V±plusmn; 10%
Power frequency
50Hz±plusmn; 2%
Power consumption
≤2500W
Temperature
5℃~35℃
For humidity
≤85%RH
host
Width 530mm x Height 480mm x Depth 470mm
column oven
Width 248mm x Height 256mm x Depth 180mm
net weight
~45kg