Purpose:
Suitable for collecting TVOC and SO in the air2、 NOX、O3、 NH3Waiting for harmful gas samples.
Adopting standards:
HJ/T 375-2007 "Environmental Air Sampler"
JJG 956-2013 "Verification Regulations for Atmospheric Samplers"
Main features:
1. Adopting constant current sampling technology to automatically control the sampling flow rate.
2. Able to digitally set and calibrate sampling flow rate.
3. It can automatically measure environmental temperature and atmospheric pressure, and convert the sampling volume into standard sampling volume.
4. There are two independent sampling gas paths, and a drying bottle with anti backflow function is connected in series in the gas path.
5. Able to perform single or dual channel, timed, quantitative, and interval sampling.
6. Comes with a lithium battery, capable of sampling without AC power supply.
7. Install 4 absorption bottle mounting holes, as well as adapter components for 10mL, 50mL porous glass plate absorption bottles and 10mL impact absorption bottles, allowing for the installation of 4 absorption bottles at will.
8 . Wide temperature LCD display screen, 240 × 128 large screen, Han display.
Technical indicators:
Parameters |
Parameter range |
resolution |
Indication error |
sample flow |
(0.1~1.0) L/min or (0.5~2.0) L/min |
0.01L/min |
Better than ± 2.5% |
ambient temperature |
(-55~99.9)℃ |
0.1℃ |
Better than ± 1 ℃ |
atmospheric pressure |
(50~140)kPa |
0.01kPa |
Better than ± 500Pa |
Pre measurement temperature |
(-55~99.9)℃ |
0.1℃ |
Better than ± 2.5 ℃ |
Pre measurement pressure |
(-30~0)kPa |
0.01kPa |
Better than ± 2.5% FS |
timing error |
1s |
Better than ± 0.1% |
|
Load capacity |
When the flow rate is 0.5L/min, the load is ≥ 18kPa; When the flow rate is 1.0L/min, the load is ≥ 10kPa. |
||
sampling mode |
Quantitative sampling: Sampling volume setting range (0~99999.9) L; Immediate Sampling: Sampling Time Setting Range (00h00min~99h59min) Timed sampling: Set the starting time range (00:00~23:59) and sampling time range (00h01min~23h59min) Interval sampling: The interval sampling frequency is (1~99) times, and the sampling time interval and sampling time can be set within 00h1min~23h59min. The sampling start time can be set arbitrarily. |
||
power consumption |
<10W |
||
noise |
≤50dB |
||
Working power supply |
Built in 12.6V/2.6Ah lithium battery |
||
Host size |
Length x Width x Height: (280 x 255 x 185) mm |
||
weight |
About 2.5kg |
||
Data recording volume |
Each gas path can store 100 sets of sampling data records, totaling 200 sets of data for users to query |
||
Working conditions:
1. Working environment temperature: (-30~50) ° C
2. Relative humidity of working environment: ≤ 85%
3. Atmospheric pressure: (80-105) kPa