CJDZ600-F2Characteristics of portable fluorine gas detector:
★Can simultaneously detect 1-6 gases, expandable to 18 gases, with unit free switching
Units available: PPM, mg/m3, Vol%, LEL%, PPHM, ppb, mg/L.
★Built in pump suction measurement, integrated with water vapor and dust filters, fast response, supports long-distance sampling
Built in water vapor and dust filters prevent sensors and instruments from being damaged by water vapor and dust, and can be used in high humidity and high dust environments;
Sampling distance greater than 10 meters, special gas path design, can be directly detected; Negative or positive pressure of 0.5 to 2 kilograms of gas has no effect on the measurement results.
★Rich human-machine interface
3.5Inch high-definition color screen, displaying real-time concentration, alarm, time, temperature, humidity, storage, communication, power, charging status, concentration unit, gas molecular formula, gas name and other information; The menu interface uses high-definition simulation graphics to display the function names of each menu.
★Large capacity data storage function, supporting multiple storage methods
Standard 100000 data storage capacity, larger capacity can be customized; Optional USB or SD card storage function, supporting real-time storage, scheduled storage, or only storing alarm concentration data; Supports viewing and deleting data locally, and can also upload data to a computer through USB, infrared communication, RS232 interface, Analyze data and store and print it using upper computer software. Support remote wireless transmission (optional function).
★Multiple communication interfaces and printing functions
Infrared communication interface (optional), USB interface, RS232 interface automatic recognition, optional built-in or external wireless infrared printer.
Print content: company name, gas name, date and time, environmental temperature and humidity, concentration data, test results (whether qualified).
★High temperature gas detection (optional)
Optional high temperature sampling and cooling filter handle or high temperature and high humidity pretreatment system for detectionSmoke at a temperature of 1300 degrees. Gas detection at higher temperatures can be customized.
★Three display modes can be switched
Simultaneously display four gas concentrations, with large font circulating to display the concentration of a single channel gas, and real-time curve display.
★Graphical display
Reflect the trend of gas concentration changes over a period of time in the form of a curve.
★Infrared remote control
Standard infrared remote control, with one click operation through the infrared remote control function keys, such as modifying alarm points, concentration calibration, zero calibration, mute, factory reset, entering menus, viewing records, 4mA output, 20mA output (this function is required for calibrating other acquisition devices).
●Data recovery function
You can choose to partially or fully restore, eliminating the worries caused by misoperation.
●Can be set to display the maximum value, minimum value, and average value
●Standard USB charging, with charging protection function, supports USB hot swapping
Can be charged with a computer or power bank, compatible with mobile phone chargers; Has overcharge, overdischarge, overvoltage, short circuit, and overheating protection; 5-level precise power display; The detector can work normally during charging.
●Adopting a 10000mA high-capacity rechargeable polymer battery, it can work continuously for a long time
●High precision temperature and humidity measurement (optional)
●Support real-time detection or scheduled detection
When not detecting, the pump can be turned off to extend the startup time.
●Multiple alarm methods, with multi-dimensional and three-dimensional indication of alarm status during alarm
Includes sound and light alarms, vibration alarms, and visual alarms on the display screen. The types of alarms include concentration alarm, undervoltage alarm, and fault alarm.
●Multiple alarm mode settings
Low alarm, high alarm, interval alarm, weighted average alarm.
●Misoperation recognition function
Automatic identification and prevention of concentration calibration errors can avoid adverse effects caused by human factors.
●Zero point automatic tracking, long-term use is not affected by zero drift
●Multi level calibration of target points to ensure measurement linearity and accuracy
●Chinese and English interface selectable
●Wide working temperature: -40~+70 ℃, supports temperature compensation
●log recording
Record calibration logs, maintenance logs, fault records, fault resolution measures, sensor life expiration reminder, and next concentration calibration time reminder function.
●Meets the requirements of intrinsic safety circuit design, anti-static, anti electromagnetic interference, and has passed national standard testing and CMC measuring instrument production license certification
●Protection level reaches IP66, waterproof, dustproof, explosion-proof, and shockproof
CJDZ600-F2Technical parameters of portable fluorine gas detector:
|
Gas detection
|
Fluorine gas/F2
Optional: Can simultaneously detect any 1-6 gas concentrations and temperature/humidity, expandable to 18 gases
|
|||||||
|
Testing Scope
|
0-1、10、20、50、200ppm
|
|||||||
|
resolution
|
0.001ppm、0.1ppm
|
|||||||
|
Detection Principle
|
electrochemistry
|
|||||||
|
Sensor lifespan
|
2-3year
|
|||||||
|
Detection method
|
Built in pump suction, flow rate 800 milliliters per minute
|
|||||||
|
display mode
|
3.5A large screen with a resolution of 320X240 inches and high-definition color display
|
|||||||
|
detection accuracy
|
≤± 2% F.S (higher precision can be customized)
|
uncertainty
|
≤±2%
|
|||||
|
Linearity
|
≤±2%
|
Repeatability
|
≤±2%
|
|||||
|
alarm method
|
Sound and light alarm, visual alarm, sound and light+visual alarm, turn off alarm
|
|||||||
|
response time
|
T90≤ 20 seconds
|
recovery time
|
≤ 30 seconds
|
|||||
|
working power supply
|
DC3.6V
|
|||||||
|
battery capacity
|
3.6VDC, 6000mA rechargeable polymer battery with overcharge, overdischarge, overvoltage, and short circuit protection functions
|
|||||||
|
Usage environment
|
Temperature -40 ℃~+70 ℃; Relative humidity ≤ 0-99% RH
|
|||||||
|
Sample gas temperature
|
-40℃~+70 ℃, optional high temperature sampling and cooling filter handle or high temperature and high humidity pretreatment system can detect flue gas concentration up to 1300 ℃
|
|||||||
|
temperature measurement
|
-40℃~+120 ℃ accuracy 0.5 ℃
|
Humidity measurement
|
0-100%RHAccuracy 3% RH
|
|||||
|
data storage
|
Standard capacity of 100000 pieces of data, optional CF card and USB storage, unlimited capacity, supports local viewing, deletion or data export,
Free upper computer communication software, storage function defaults to off state, can be set to on state, storage time interval can be set arbitrarily
|
|||||||
|
communication interface
|
Infrared, USB, RS232 automatic recognition
|
|||||||
|
printer
|
Optional, built-in micro printer or external micro wireless infrared printer
|
|||||||
|
wireless transmission
|
Optional, data can be wirelessly transmitted to monitoring devices such as mobile phones, remote monitoring centers, and monitoring computers. The upper computer can be used for data analysis, storage, printing, and other functions on the computer
|
|||||||
|
Interface language
|
Chinese or English can be set, default Chinese interface
|
|||||||
|
Explosion proof type
|
intrinsically safe
|
Explosion-proof marking
|
Explosion proof Exd Ⅱ CT6
|
|||||
|
Protection level
|
IP66Rain and water splashes, dust prevention
|
|||||||
|
Overall dimensions
|
230×210×120mm(L×H×W)
|
weight
|
2.5Kg
|
|||||
|
Standard Appendix
|
Instruction manual, certificate of conformity, warranty card, USB charger (including data cable), high-end aluminum alloy instrument box, CD (upper computer communication software), 40cm stainless steel sampling handle (1-meter hose)
|
|||||||
|
Optional accessories
|
Built in micro printer, external micro wireless infrared printer, temperature and humidity measurement function, USB storage, wireless data communication function, high temperature sampling and cooling filter handle, high temperature and high humidity pre-processing system, external remote sampling pump, multiple humidity and dust filters, 1.2m extendable sampling handle (1-meter hose)
|
|||||||
|
Application scenarios
|
All occasions requiring gas concentration detection, such as petroleum, chemical, pharmaceutical, environmental protection, flue gas analysis, air treatment, scientific research institutions, etc
|
|||||||