1、 Product Introduction
The CIC-D100 ion chromatograph is a structurally simple and highly stable ion chromatograph that maintains the consistent economic and durable characteristics of a five pole circuit. It is equipped with an independently developed observatory intelligent workstation, which can not only achieve intelligent control of various instrument functional components, but also has powerful data processing functions. Relying on internationally leading chromatographic column technology, it can analyze anions and cations, cyanide ions, iodide ions, sugars, small molecule organic acids, etc. by configuring conductivity detectors and amperometric detection. It is widely applicable in fields such as environmental protection, disease control, chemical industry, hydrology, mining and metallurgy.
2、 Product parameters
parts
technical parameter
host
CIC-D100
Infusion pump
model: SH-100A(D100)
Maximum pressure resistance: 42MPa (PEEK material); Flow range: 0.001-9.999mL/min
Allowable error of flow setting value: (0.2-0.5) mL/min, ± 5%;
(0.5-1.0)mL/min,±3%;
Greater than 1.0mL/min, ± 2%
Traffic stability: (0.2-0.5)mL/min≤3%;
(0.5-1.0)mL/min≤2%;
Greater than 1.0mL/min ≤ 2%Six way valve
Manual valve: VI-11 (C series D100)
Maximum pressure resistance: 35MPa; Tube outer diameter: 1/16 "
Rotor contact material: PEEK;’ Dielectric contact material: PEEK/ceramicConductivity cell
Five pole constant temperature conductivity cell: SHD-3
Type of conductivity cell: Five pole circular passivated 316 stainless steel; Pool volume: ≤ 0.8μL
Detection range: 0-35000 μ S/cm; Detection resolution: 0.0020nS/cm
Output voltage:- 6000~+6000mV adjustable; Baseline noise: ≤ 0.05%FS
Baseline drift: ≤ 3%FS/30min; Constant temperature accuracy: 5-60 ± 0.01 ℃
Maximum withstand voltage: 10MPa; Temperature compensation: 1.7%/℃
Minimum detection limit: Cl-≤0.0005μg/mL
Li+≤0.001μg/mL
BrO3-≤0.001μg/mL
Instrument linearity: ≥ 0.999; Qualitative repeatability: ≤ 1.0%; Quantitative repeatability: ≤ 1%Suppressor
Anion inhibition function: compatible with carbonate, hydroxide systems and other eluents
Advanced continuous self regenerating micro film electrochemical suppression technology, without the need for additional regeneration fluid;
High inhibitory capacity, low background conductivity;
Low noise, good stability;
High pressure resistance, no leakage at up to 6MPa, normal operation at 2MPa;
Smaller dead volume, more sensitive response signal;Cation inhibition function: compatible with leaching solutions such as nitric acid and methane sulfonic acid systems
Advanced continuous self regenerating micro film electrochemical suppression technology, without the need for additional regeneration fluid;
High inhibitory capacity, low background conductivity;
Low noise, good stability;
High pressure resistance, no leakage at up to 6MPa, normal operation at 2MPa;
Smaller dead volume, more sensitive response signal;Column temperature box
Heating method: Embedded heat conduction constant temperature column temperature box
Temperature control range: 20-60 ℃; Allowable error of temperature setting value for column temperature box: ± 2 ℃
Temperature control accuracy: ± 0.01 ℃; Temperature stability: 1 ℃/h
3、 Product Features
1. Embedded heat conduction heating technology, with stable temperature and fast speed, ensures accurate and reliable experimental data;
2. Temperature controlled five pole conductivity detector with intelligent temperature control to ensure better analysis accuracy;
3. Self regenerating electrolytic biofilm suppressor, high voltage resistance, small dead volume, and corresponding high signal;
4. Five pole circuit system, mature and stable technology, and stable experimental data;
5. Observatory intelligent workstation, integrated control, compatible with multiple instruments, customized graphics.
Juchuang Environmental Protection provides you with comprehensive technical support and perfect after-sales service! Details consultation:
-
1. Embedded heat conduction heating technology, with stable temperature and fast speed, ensures accurate and reliable experimental data;
2. Temperature controlled five pole conductivity detector with intelligent temperature control to ensure better analysis accuracy;
3. Self regenerating electrolytic biofilm suppressor, high voltage resistance, small dead volume, and corresponding high signal;
4. Five pole circuit system, mature and stable technology, and stable experimental data;
5. Observatory intelligent workstation, integrated control, compatible with multiple instruments, customized graphics.
-
parts
technical parameter
host
CIC-D100
Infusion pump
model: SH-100A(D100)
Maximum pressure resistance: 42MPa (PEEK material); Flow range: 0.001-9.999mL/min
Allowable error of flow setting value: (0.2-0.5) mL/min, ± 5%;
(0.5-1.0)mL/min,±3%;
Greater than 1.0mL/min, ± 2%
Traffic stability: (0.2-0.5)mL/min≤3%;
(0.5-1.0)mL/min≤2%;
Greater than 1.0mL/min ≤ 2%Six way valve
Manual valve: VI-11 (C series D100)
Maximum pressure resistance: 35MPa; Tube outer diameter: 1/16 "
Rotor contact material: PEEK;’ Dielectric contact material: PEEK/ceramicConductivity cell
Five pole constant temperature conductivity cell: SHD-3
Type of conductivity cell: Five pole circular passivated 316 stainless steel; Pool volume: ≤ 0.8μL
Detection range: 0-35000 μ S/cm; Detection resolution: 0.0020nS/cm
Output voltage:- 6000~+6000mV adjustable; Baseline noise: ≤ 0.05%FS
Baseline drift: ≤ 3%FS/30min; Constant temperature accuracy: 5-60 ± 0.01 ℃
Maximum withstand voltage: 10MPa; Temperature compensation: 1.7%/℃
Minimum detection limit: Cl-≤0.0005μg/mL
Li+≤0.001μg/mL
BrO3-≤0.001μg/mL
Instrument linearity: ≥ 0.999; Qualitative repeatability: ≤ 1.0%; Quantitative repeatability: ≤ 1%Suppressor
Anion inhibition function: compatible with carbonate, hydroxide systems and other eluents
Advanced continuous self regenerating micro film electrochemical suppression technology, without the need for additional regeneration fluid;
High inhibitory capacity, low background conductivity;
Low noise, good stability;
High pressure resistance, no leakage at up to 6MPa, normal operation at 2MPa;
Smaller dead volume, more sensitive response signal;Cation inhibition function: compatible with leaching solutions such as nitric acid and methane sulfonic acid systems
Advanced continuous self regenerating micro film electrochemical suppression technology, without the need for additional regeneration fluid;
High inhibitory capacity, low background conductivity;
Low noise, good stability;
High pressure resistance, no leakage at up to 6MPa, normal operation at 2MPa;
Smaller dead volume, more sensitive response signal;Column temperature box
Heating method: Embedded heat conduction constant temperature column temperature box
Temperature control range: 20-60 ℃; Allowable error of temperature setting value for column temperature box: ± 2 ℃
Temperature control accuracy: ± 0.01 ℃; Temperature stability: 1 ℃/h