Product Usage: The Xenon Arc Lamp Aging Test Chamber uses a xenon arc lamp that can simulate the full spectrum of sunlight to reproduce destructive light waves present in different environments. It can provide corresponding environmental simulation and accelerated testing for scientific research, product development, and quality control.
1、 Technical parameters:
1. Temperature range: RT+10 ℃~70 ℃
2. Humidity range: 50% to 98% R% H
3. Blackboard temperature: 63 ℃~ 100 ℃± 3 ℃
4. Temperature fluctuation: ≤± 0.5 ℃
5. Temperature uniformity: ≤ 2 ℃
6. Humidity fluctuation:+2, -3% R% H
7. Glass filter: 1 piece
8. Xenon lamp source: air-cooled lamp tube
9. Number of lamp tubes: 2 (one for backup)
10. Rainfall duration: 0-9999 minutes, adjustable for continuous rainfall
11. Rainfall cycle: 0-240min, adjustable interval (intermittent) rainfall
12. Spray cycle: 18min/102min or 12min/48min (spray time/no spray time)
13. Rainwater pressure: 0.12-0.15Mpa
14. Spray nozzle aperture: Ø 0.8mm
15. Xenon lamp power: 2.5KW
16. Heating power: 1.5KW
17. Humidification power: 1.5KW
18. Sample tray size: 320 × 450mm
19. Distance between sample holder and lamp: 230-280mm
20. Spectral wavelength: 290nm to 800nm
21. Continuous adjustable light cycle time: 0-999 H, M, S
22. Irradiance (adjustable): The average irradiance intensity is 550W/㎡± 10% when the spectral wavelength is between 290nm and 800nm
23. Total power: 6.0KW
24. Power requirements: AC380 (± 10%) V/50HZ three-phase five wire system
IIBeijing BD/SN-L Xenon Arc Lamp Aging Testing MachineBox material:
1. The outer shell is made of high-quality (t=1.2mm) A3 steel plate CNC machine tool, and the surface of the shell is sprayed with plastic for a smoother and more beautiful appearance;
2. The inner liner is made of imported (SUS) 304 high-quality stainless steel plate;
3. The insulation material adopts high-density glass fiber cotton and high-pressure polyurethane foam insulation;
4. The mixing system adopts a long axis fan motor and a stainless steel multi blade impeller that is resistant to high and low temperatures to achieve strong convection and vertical diffusion circulation;
5. Double layer high tensile sealing strips with high temperature resistance are used between the door and the box to ensure the sealing of the testing area;
6. Using non reactive door handles makes operation easier;
7. The bottom of the machine adopts high-quality fixed PU movable wheels;
3、 Heating system:
1. Adopting far-infrared stainless steel high-speed heating electric heater;
2. High temperature, humidity, and lighting are completely independent systems (without interfering with each other);
3. The output power of temperature and humidity control is calculated by a microcomputer to achieve high-precision and high-efficiency electricity efficiency;
4、 Humidification system:
1. The built-in boiler steam humidifier has energy-saving and consumption reducing functions, which can save 70% of energy consumption;
2. Equipped with automatic water level compensation and water shortage alarm system;
3. The humidity control adopts P.I.D+S.S.R, and the system is coordinated and controlled in the same channel;
4. Dehumidification system: using evaporator coil dew point temperature laminar contact dehumidification method;
5、 Control system:
1. The temperature, humidity, and irradiance control instrument adopts a fully imported 7.0-inch true color touch screen controller
2. Accuracy: 0.1 ℃ (display range);
3. Resolution: ± 0.1 ℃;
4. Temperature sensor: PT100 platinum resistance thermometer;
5. Control mode: thermal balance temperature and humidity control mode; All electrical appliances use Schneider series products;
6. The temperature and humidity control adopts the P.I.D+S.S.R system for coordinated control of the same channel;
7. It has the function of automatic calculation, which can immediately correct the temperature and humidity change conditions, making temperature and humidity control more accurate and stable;
8. The circulation system adopts a single cycle, centrifugal fan, and split flow air duct structure;
6、 Refrigeration and dehumidification system control:
1. Compressor: Fully enclosed French Taikang unit;
2. Refrigeration method: single machine compression refrigeration;
3. Condensation method: air-cooled;
4. Refrigerant: R404A (environmentally friendly)
5. The entire system pipeline undergoes a 48 hour leak detection test with ventilation and pressurization;
6. The heating and cooling systems are completely independent;
7. Internal spiral refrigerant copper tube;
8. Fin slope evaporator (with automatic defrosting system)
9. Dehumidification system: using evaporator coil dew point temperature laminar contact dehumidification method;
7、 Lighting system:
1. Blackboard thermometer: metal blackboard thermometer;
2. Xenon lamp tube: The lifespan of the tube is 1800-2000 hours;
3. Irradiance control: Automatically adjust irradiance, and the panel directly displays the current lamp irradiation intensity;
8、 Spray water circulation system:
1. Water quality: Deionized water (solid content less than 20ppm)
2. Equipped with a water level display for the storage tank;
3. The sprayed water has recyclability;
4. The spray pressure can be adjusted between 0.12 and 0.15 Mpa;
5. Two nozzles were installed at the top of the studio;
9、 Protection system:
1. Fan overheat protection
2. Overall equipment under phase/reverse phase protection
3. Overload protection for refrigeration system
4. Refrigeration system overpressure protection
5. Overtemperature protection
6. Overheating of water pump, overcurrent protection
7. Other issues include electrical leakage
8. Water shortage indication
9. Automatic shutdown after fault alarm
10、 Installation site and usage conditions:
1. The installation location should take into account the heat dissipation and easy to inspect and repair position
2. The distance between the wall and any other machine should be at least 600mm.
3. Set on a flat and vibration free ground.
4. Choose good ventilation and avoid direct sunlight exposure.
5. This machine should be kept away from heat sources and flammable and explosive materials.
6. The power supply lines and water supply and drainage pipes should be shortened as much as possible.
7. Please choose a location with less dust and moisture and good ventilation for installation
8. Environmental temperature: 5 ℃~+28 ℃ (average temperature within 24 hours ≤ 28 ℃)
9. Environmental humidity: ≤ 85% R.H
PS: The operating environment needs to be below room temperature of 28 degrees Celsius and well ventilated;
The machine should be placed 80 centimeters in front, back, left, and right, and nothing should be placed there;
11、 Execution standards: Design and manufacture strictly in accordance with relevant standards such as GB/T16422., GB/T1865, GB/T9344, GB/T2423.24, etc;
TwelveBeijing BD/SN-L Xenon Arc Lamp Aging Testing MachineService commitment: 18 months warranty, free home delivery. After the installation and commissioning of the equipment is completed, relevant technical personnel will receive free operation training at the user's site, with no limit on the number of people.