Brief introduction
The AN-WQSTU turbidity sensor is based on the combined infrared absorption scattering light method and can continuously and accurately measure turbidity using the ISO7027 method. According to ISO7027 infrared double scattering light technology, turbidity values are determined without being affected by chromaticity. According to the usage environment, it can be optionally equipped with self-cleaning function. Stable data and reliable performance; Built in self diagnosis function to ensure accurate data; Installation and calibration are simple.
It can be widely used for turbidity monitoring in sewage treatment plants, water treatment plants, water stations, surface water, industrial and other fields.
2、 Working principle
The turbidity sensor is equipped with an infrared tube packaged with IR958 and PT958. When light passes through a certain amount of water, the amount of light transmitted depends on the degree of water pollution. The more polluted the water, the less light passes through. The light receiving end converts the transmitted light intensity into the corresponding current size. The more transmitted light, the higher the current, and vice versa. By measuring the current at the receiving end, the degree of water pollution can be calculated.
3、 Advantages and Characteristics
Infrared scattered light technology can eliminate the influence of sample color
Optional cleaning brush with automatic cleaning function, greatly reducing sensor maintenance
Digital sensors with strong anti-interference ability and long transmission distance
Standard digital signal output, which can be integrated and networked with other devices without a controller
The on-site installation of sensors is convenient and fast, achieving plug and play functionality
Sensor power supply positive and negative polarity reverse protection
Sensor RS485 A/B terminal misconnected power protection
Optional wireless data transmission module
Long operating cycle, low power consumption
4、 Technical parameters
Measurement range: 0.01-100 NTU, 0.01-4000 NTU (range can be customized)
Display accuracy: less than ± 2% of the measured value or ± 0.1NTU, whichever is greater
Output: RS485 MODBUS
Operating temperature: 0-40 ℃ (non freezing)
Storage temperature: -15~60 ℃
Pressure range: ≤ 0.4Mpa
Flow velocity: ≤ 2.5m/s, 8.2ft/s
Calibration: Sample calibration, slope calibration
Power supply: 12VDC
Size: Diameter 60mm * Length 256mm
Weight: 1.65KG
Cable: Standard 10 meters
Material: Main body: SUS316L (regular version), titanium alloy (seawater version) Upper and lower covers: PVC, cable: PVC
Warranty: 1 year